NMR Spectroscopy Basics
Nuclear Magnetic Resonance Spectroscopy
What NMR spectroscopy measures
Nuclear magnetic resonance (NMR) spectroscopy examines certain atomic nuclei by placing them in a magnetic field and applying radio-frequency electromagnetic radiation. The nuclei absorb energy and change their spin state; as they return to a lower-energy state, the instrument detects a signal.
In organic chemistry, the nuclei most often studied are:
Protons—the nuclei of ordinary hydrogen atoms.
Carbon-13 nuclei—an isotope of carbon with a nucleus that can be detected by NMR.
The frequencies and patterns of the signals provide information about the number and chemical environments of these nuclei in a molecule.
Nuclear spin and magnetic behavior
A nucleus with nonzero spin behaves like a tiny magnet because its spin gives it a magnetic moment. A proton has a spin quantum number of I=21, so it can interact with an external magnetic field.
The number of possible nuclear spin states is determined by:
For a proton, I=21, so:
Thus, a proton has two possible spin states in a magnetic field.
The exact spin behavior depends on the nucleus. A useful introductory rule is that nuclei with an odd number of protons or an odd number of neutrons often have nonzero spin, while nuclei with both even proton and neutron numbers commonly have zero spin. The key requirement for NMR is that the nucleus being observed has nonzero spin.
Energy levels in a magnetic field
Alignment and energy
When no external magnetic field is present, the two proton spin states have the same energy. Applying a magnetic field, written B0, separates them into two energy levels:
The alpha state is the lower-energy state, commonly described as aligned with the applied field.
The beta state is the higher-energy state, commonly described as opposed to the applied field.
The energy difference between these states is:
Here, h is Planck’s constant and ν is the frequency of the radiation that corresponds to the energy difference.
At ordinary thermal conditions, slightly more nuclei occupy the lower-energy alpha state than the higher-energy beta state. Radio-frequency energy can promote some nuclei from alpha to beta. When they return to alpha, the instrument detects a signal.
Effect of magnetic-field strength
The energy gap between the states increases as the applied magnetic field becomes stronger. Consequently, stronger fields require higher-frequency radiation to induce the transition.
This relationship is expressed as:
where γ is the gyromagnetic ratio, a constant that depends on the type of nucleus, and B0 is the applied magnetic field. Protons and carbon-13 nuclei have different gyromagnetic ratios, so they respond at different frequencies under the same magnetic-field conditions.
Resonance and signal detection
Resonance occurs when a nucleus absorbs radio-frequency energy whose frequency matches the energy gap between its spin states. The absorption promotes the nucleus from the lower-energy alpha state to the higher-energy beta state.
A simplified sequence is:
A sample is placed in a strong magnetic field, causing its nuclei to occupy the two spin states, with a slight excess in the lower-energy state.
The instrument applies a radio-frequency pulse.
Some nuclei absorb energy and change spin state.
After the pulse, nuclei relax toward the lower-energy state.
The instrument detects the resulting signal.
NMR uses radio-frequency radiation, which is relatively low in energy. The source describes typical operating frequencies ranging from tens to hundreds of megahertz, depending on the instrument and nucleus being measured.
The NMR instrument
An NMR instrument places the sample in a strong, carefully controlled magnetic field. A narrow sample tube is positioned in a probe within the magnet’s central bore. The probe delivers radio-frequency pulses to the sample and detects the resulting signal.
Many high-field instruments use superconducting magnets. These magnets are cooled to very low temperatures to reduce electrical resistance and sustain powerful magnetic fields. A typical design includes:
A central region containing the superconducting magnet.
A very cold helium region, around 4 K.
A surrounding liquid-nitrogen region, around 77 K.
An outer vacuum chamber that helps limit heat transfer.
A uniform field is important: if different parts of the sample experience substantially different fields, their signals become less precise. The instrument therefore adjusts the field to make it as uniform as possible; this process is called shimming.
Why high-field instruments are useful
A stronger magnetic field increases the energy separation between the spin states and raises the resonance frequency. This generally improves signal strength and helps distinguish closely spaced signals. High-field instruments are especially useful for complex samples, which can produce many closely spaced peaks. Smaller organic molecules can often be studied with lower-field instruments.
From the detected signal to a spectrum
After a radio-frequency pulse, the nuclei produce a time-varying signal called the free induction decay (FID). The FID is recorded in the time domain: it shows how the detected signal changes over time as the nuclear response decays.
The instrument applies a Fourier transform to convert the FID into a frequency-domain spectrum. The resulting spectrum displays signals at frequencies associated with the nuclei in the sample.
The measured response involves the nuclei’s behavior in the magnetic field and the instrument’s detection of the changing signal. In introductory terms, nuclei absorb energy during excitation and produce a detectable signal as they relax.
Why molecules produce multiple signals
Not all nuclei of the same type have the same surroundings. NMR is sensitive to a nucleus’s chemical environment, so nuclei in different environments generally resonate at different frequencies and produce separate signals.
Chemically equivalent nuclei
Nuclei that are equivalent because of a molecule’s symmetry or identical surroundings produce the same signal. For example, if several hydrogens in a molecule are related by symmetry and experience the same environment, their signals appear together as one peak.
Distinct environments
Consider a carbon chain with a chlorine atom at one end, such as ClCH2CH2CH3. Its hydrogens fall into three broad environments:
The two hydrogens on the carbon nearest chlorine.
The two hydrogens on the middle carbon.
The three hydrogens on the terminal carbon.
These environments are different, so the molecule produces three signals. Their relative proton counts are 2:2:3.
The electron distribution contributes to these differences. Electrons also have magnetic properties and influence the local magnetic field experienced by a nucleus. An electron-withdrawing group such as chlorine pulls electron density toward itself, changing the environments of nearby hydrogens. The effect is usually strongest for nuclei closest to the group and decreases with distance.
Information provided by an NMR spectrum
A proton NMR spectrum can provide several kinds of structural information.
1. Number of distinct proton environments
The number of signals indicates how many different kinds of proton environments are present, assuming the signals can be resolved. A single signal may represent several equivalent hydrogens; it does not necessarily mean the molecule contains only one hydrogen.
2. Chemical shifts
The chemical shift describes the position of a signal relative to a reference. It reflects the nucleus’s local electronic environment. Electron-withdrawing groups commonly cause nearby hydrogens to appear at larger chemical shifts, while less electron-poor environments tend to appear at smaller shifts.
3. Relative numbers of protons
The integration of a signal measures its area and indicates the relative number of hydrogens contributing to it. For example, two signals with integral values in a 1:3 ratio represent proton groups in that relative ratio. For a molecule with groups containing two, two, and three hydrogens, the integrals should be in a 2:2:3 ratio.
Integration gives a ratio, not automatically an absolute count. The ratio is interpreted together with the molecular formula and other structural information.
4. Spin–spin splitting
Signals can be split into multiple nearby peaks through interactions between the spins of neighboring nuclei. This is called spin–spin splitting or spin–spin coupling. The resulting pattern can provide information about nearby protons. It is a separate source of structural information from signal count, chemical shift, and integration.
Together, these features help identify how atoms and groups are arranged in a molecule.
Chemical shift and the TMS reference
Tetramethylsilane
Proton NMR commonly uses tetramethylsilane (TMS) as a reference. Its hydrogens are assigned a chemical shift of zero. Other signals are measured relative to the TMS signal.
A direct frequency difference from the reference can be written as:
However, this difference in hertz depends on the operating frequency of the spectrometer. The same sample can therefore have different frequency separations in hertz on different instruments.
Chemical shift in parts per million
To make measurements comparable across instruments, the frequency difference is divided by the spectrometer’s operating frequency. The chemical shift is expressed in parts per million:
When the numerator and denominator are expressed in matching frequency units, their units cancel, leaving a dimensionless value. The result is conventionally reported in ppm.
For example, a signal 436 Hz from TMS on a 60 MHz instrument has a chemical shift of about 7.27 ppm. On a 300 MHz instrument, the same sample’s frequency separation is about 2181 Hz, but the chemical shift remains about 7.27 ppm:
This normalization allows chemical shifts to be compared between instruments with different operating frequencies.
Signal intensity and integration
The vertical axis of an NMR spectrum represents signal intensity. The area under a signal, rather than its peak height alone, is used for integration. Integrals provide the relative amounts of the nuclei contributing to the signals.
For example, if two integrated signals have areas in the ratio 3:2, the corresponding proton groups contain hydrogens in a relative ratio of three to two. A spectrum therefore reveals both how many distinct proton environments are present and how many protons belong to each environment relative to the others.
NMR energy compared with other molecular transitions
NMR transitions involve small energy differences. The source gives an approximate energy scale of 10−6 kcalmol−1 for an NMR transition, compared with roughly 10−4 kcalmol−1 for molecular rotation and approximately 1 to 100 kcalmol−1 for molecular vibrations.
These are broad comparisons: NMR probes nuclear spin transitions, while vibrational spectroscopy probes changes in molecular vibrations, including changes in bond lengths. The different energy scales mean that the methods reveal different aspects of molecular structure.
Key principles
NMR detects nuclei with nonzero spin in a magnetic field.
A proton has two spin states, with a lower-energy alpha state and a higher-energy beta state.
Radio-frequency energy matching the energy gap causes resonance.
The magnetic-field strength and the nucleus’s gyromagnetic ratio determine its resonance frequency.
Distinct nuclear environments produce distinct signals.
Signal position gives chemical shift, signal area gives relative proton count, and signal splitting gives information about neighboring nuclei.
TMS provides a zero reference, and chemical shifts are reported in ppm so results can be compared across different spectrometers.