MRI Physics Principles

Magnetic Resonance Imaging (MRI): Principles and Physics

Introduction to Nuclear Magnetic Resonance (NMR)

MRI is based on the physics principle of nuclear magnetic resonance (NMR), first described in the 1930s and 1940s. NMR involves the interaction of certain atomic nuclei, radio frequency (RF) energy, and a strong magnetic field.

Basic Concepts: The Electromagnetic Spectrum

Radio frequency (RF) energy is part of the electromagnetic spectrum, which also includes visible light and X-rays. RF waves lie at the far left (low frequency, long wavelength), visible light in the middle, and X-rays on the right (high frequency, short wavelength).

All electromagnetic waves are defined by:

  • Wavelength: the distance between successive peaks of the wave

  • Frequency: the number of cycles the wave completes per second (measured in Hz)

  • Amplitude: the height or power of the wave

Phase and Frequency

When describing two or more waves, phase becomes important. Two waves with the same frequency that line up completely are in phase. If one wave is shifted horizontally (delayed), they become out of phase or phase shifted. Waves that start at the same time but have different frequencies are initially in phase but will drift apart over time.

In NMR and MRI, radio frequency signals are described by their unique phase and frequency — these properties allow spatial localization of signals.

Image Elements: Pixels and Voxels

A pixel (picture element) is a single point on a 2D image. MRI images are composed of voxels (volume elements) — each square on the picture corresponds to a three-dimensional volume of tissue in the body. The MRI machine measures the NMR signal from each voxel, localizes it in 3D space, and plots the intensities on a matrix (commonly 256×256256 \times 256256×256 or 512×512512 \times 512512×512) to form a visible image.


Electromagnetism and the Hydrogen Atom

Electromagnetism Basics

A flowing current of electrons generates an associated magnetic field oriented 90° from the direction of current flow. Coiling the wire concentrates the magnetic field. The direction of the field is given by the right-hand rule: curl the fingers of the right hand in the direction of current flow; the thumb points in the direction of the magnetic field.

The Hydrogen Atom as an NMR Probe

The hydrogen atom consists of a single proton in the nucleus with a single orbiting electron. For NMR, the electron is of little interest — the focus is on the single central proton (hence "nuclear" in NMR). The proton does not sit statically; it rotates on its axis, similar to planets and stars. Because the proton is positively charged, its rotation generates a small magnetic moment (a tiny magnet). The direction of this magnetic field is again given by the right-hand rule: curl fingers in the direction of rotation, thumb points along the magnetic field.

These individual proton magnets are often called spins (a quantum mechanical property, but the classical spinning charge model is useful for understanding).

Abundance of Hydrogen in the Body

Hydrogen is plentiful in the human body, making it ideal for imaging:

  • An adult is about 60% water (H₂O) — two hydrogen atoms per oxygen

  • Energy-storing molecules like fat (triglycerides) and carbohydrates (e.g., glucose, C₆H₁₂O₆) contain abundant hydrogen

  • Free fatty acids have 2–3 hydrogens per carbon atom in their carbon chains


Behavior of Protons in a Magnetic Field

Random Orientation Without a Field

While walking around on Earth, the proton magnets are randomly oriented, canceling each other out. The body is not normally magnetic.

Alignment in a Strong Magnetic Field

When placed inside a strong magnetic field (like an MRI scanner), the proton spins align with the main magnetic field (B₀):

  • Most align with the field (low energy state)

  • A few align directly opposite (high energy state)

The orientation depends on the energy of each proton. Protons with a little extra energy (e.g., from local heat) align against the field — these are in the high energy state. Protons with less energy align with the field — these are in the low energy state.

Key concept: The distribution between high and low energy states follows the Boltzmann distribution. The net magnetic effect is the vector sum of all spins.

Precession and Larmor Frequency

The proton spins do not simply point statically — they precess (wobble) around the direction of the main magnetic field, much like a spinning top as it slows down. The rate of precession is given by the Larmor equation:

ω0=γB0\omega_0 = \gamma B_0ω0​=γB0​

where ω0\omega_0ω0​ is the Larmor (angular) frequency, γ\gammaγ is the gyromagnetic ratio for hydrogen (42.58 MHz/T), and B0B_0B0​ is the magnetic field strength.

For hydrogen at common field strengths:

  • 1 Tesla: Larmor frequency = 42.5 MHz

  • 1.5 Tesla: 63.75 MHz (1.5 × 42.5)

  • 3 Tesla: 127.5 MHz (3 × 42.5)

This frequency is in the radio frequency range and is critically important — it determines the exact RF frequency needed to excite the spins.

Net Magnetization and Longitudinal Magnetization

If all precessing protons are moved to a common origin on a 3D graph, most point in the direction of B₀ (low energy) with a few pointing opposite (high energy). The vector sum of all these magnetic moments produces a net magnetization vector pointing along B₀. This is called the longitudinal magnetization (M_z) because it is parallel to the main magnetic field.

Longitudinal magnetization cannot be measured directly — it is a static vector aligned with B₀ and does not induce a current in a receiver coil. However, it can be changed by applying RF energy.


The RF Pulse: Resonance and Transverse Magnetization

Applying a Resonant RF Pulse

If a radio frequency pulse of exactly the Larmor frequency (e.g., 42.5 MHz at 1 T) is transmitted into the tissue, two things happen:

  1. Energy absorption: Protons absorb the RF energy and some flip from the low energy state to the high energy state. If enough energy is applied to equalize the populations (say 50% in each state), the longitudinal magnetization reduces to zero — the opposing magnetic forces cancel.

  2. Phase synchronization: The RF pulse forces the precessing protons to synchronize and spin together (move in phase). This is the resonance part of NMR — the protons resonate at the same frequency as the applied RF.

Formation of Transverse Magnetization

When the spins are in phase and precessing together, their individual magnetic moments add up to produce a net magnetic force oriented horizontally — 90° to the longitudinal direction. This is called transverse magnetization (M_{xy}).

The transverse magnetization rotates at the Larmor frequency and can be detected by a receiver coil.

Detection: Electromagnetic Induction

Just as a current can create a magnetic field, a moving magnetic field can induce a current in a coil. This is the principle of a generator. If a magnet is moved through a coil of wire, a current is generated. If the magnet spins, a sinusoidal (alternating) current is produced.

Similarly, the rotating transverse magnetization (a rotating magnetic vector) induces a small but measurable radio frequency signal in a nearby receiver coil. This signal is the NMR signal — the fundamental measurement in MRI.


Relaxation: T2 and T1

After the RF pulse is turned off, the protons gradually return to their equilibrium state through two independent relaxation processes:

T2 Relaxation (Spin-Spin Relaxation)

Protons are all positively charged and repel each other. Once the RF pulse is removed, they begin to dephase — they spread apart due to local magnetic field inhomogeneities and spin-spin interactions. As they lose phase coherence, the transverse magnetization decays.

  • No net energy transfer to the surroundings; energy is exchanged among the spins.

  • The time constant for this decay is called T2.

  • T2 relaxation destroys the transverse magnetization.

T1 Relaxation (Spin-Lattice Relaxation)

High-energy protons spontaneously return to the low energy state, releasing the previously absorbed energy into the surrounding tissues as heat. This involves an actual energy transfer from the spins (the "spin system") to the surrounding molecular environment (the "lattice").

  • As protons fall back to baseline, longitudinal magnetization regrows.

  • The time constant for this recovery is called T1.

  • T1 relaxation restores the equilibrium longitudinal magnetization.

Summary of the NMR Cycle

A complete NMR excitation and detection cycle:

  1. RF pulse tuned to Larmor frequency flips 50% of spins into high energy state and synchronizes their precession, rotating longitudinal magnetization into the transverse plane.

  2. Transverse magnetization precesses and induces a detectable RF signal in the receiver coil.

  3. T2 decay: Spins dephase, transverse magnetization decays (signal fades).

  4. T1 recovery: Spins return to low energy state, longitudinal magnetization regrows, and the absorbed energy is dissipated as heat.


Tissue Contrast: T1 and T2 Differences

Protons in different tissues experience different local environments — some are in free water, others bound in proteins or fat. These differences affect T1 and T2 relaxation times. The key is that fat and water have characteristic relaxation differences.

Tissue

Proton environment

T2 (dephase rate)

T1 (recovery rate)

Fat

Protons in fatty acids are relatively fixed in position.

T2 decay is fast — spins dephase quickly.

T1 recovery is fast — energy is given up rapidly to surrounding lattice.

Water

Protons in free water move freely.

T2 decay is slow — spins remain in phase longer.

T1 recovery is slow — energy is held longer, protons stay in high energy state.

These differences can be accentuated by carefully choosing the timing of RF pulses and signal acquisition.


Pulse Sequences: TR and TE

Two timing parameters define the pulse sequence:

  • Repetition Time (TR): The time between successive RF pulses.

  • Echo Time (TE): The time between the RF pulse and when the signal (echo) is measured.

By selecting TR and TE, different weightings of T1, T2, and proton density are achieved.

T2-Weighted Image (T2W)

To emphasize differences in T2 relaxation:

  • Long TR: Allows nearly complete T1 recovery, minimizing T1 effects.

  • Long TE: Allows time for T2 decay to separate tissues — fast T2 tissues (fat) lose signal, slow T2 tissues (water) retain strong signal.

Result: Water appears bright (white), fat appears dark (gray/black) on a conventional T2-weighted image.

T1-Weighted Image (T1W)

To emphasize differences in T1 relaxation:

  • Short TR: RF pulses are applied rapidly, so tissues with slow T1 recovery (water) are still in high energy state — they are "saturated" and produce little signal when hit again.

  • Short TE: Minimizes T2 effects — little time for dephasing.

Result: Fat (fast T1 recovery) gives a strong signal and appears bright; water (slow T1 recovery) appears dark.

Proton Density (PD) Weighted Image

To minimize both T1 and T2 effects and show the relative number of protons per voxel:

  • Long TR (minimizes T1 effects) + Short TE (minimizes T2 effects)

Result: Signal intensity reflects the density of hydrogen protons in the tissue.

Summary: T1-weighted: short TR, short TE. T2-weighted: long TR, long TE. Proton density: long TR, short TE.


Historical Context

Raymond Damadian (1971)

Dr. Raymond Damadian published "Tumor Detection by Nuclear Magnetic Resonance" in Science (March 1971). He proposed that NMR signals from tumors could be differentiated from normal tissues and detected. His original concept was a whole-body NMR detector — a screening tool to detect malignancy by characteristic T1 and T2 signals, not an anatomic imager. While his tumor detector never became a clinical reality, MR spectroscopy (a related technique) is used today alongside standard MRI to help characterize tissues (e.g., distinguishing benign from malignant processes).

Paul Lauterbur (1973)

Dr. Paul Lauterbur published "Image Formation by Induced Local Interactions: Examples Employing Nuclear Magnetic Resonance" in Nature (March 1973). He described a technique using magnetic field gradients to localize NMR signals. In his experiment, he used two 1 mm capillary tubes filled with water (H₂O), submerged in a larger tube of heavy water (D₂O). Deuterium (an isotope of hydrogen with one proton and one neutron) has a different NMR behavior and gives a weak signal, so it acted as a "silent" background. Using gradients, Lauterbur created a primitive 2D image of the two water tubes — the first MRI image. He called his technique zeugmatography (from Greek zeugma, meaning "that which is used for joining").


How an MRI Image Is Formed: Spatial Localization

The Main Magnetic Field and Gradients

The MRI scanner contains:

  • A superconducting magnet producing a strong, homogeneous main field (B₀) — common strengths: 1 T, 1.5 T, 3 T.

  • Three sets of gradient magnets (electromagnets) that create linear variations in the magnetic field along three orthogonal axes: Z (head-to-foot), X (left-right), Y (anterior-posterior).

Since the Larmor frequency depends on local magnetic field strength, changing the field with gradients changes the resonance frequency of protons at different spatial locations.

Slice Selection (Z-Gradient)

To select a specific slice through the body, the Z-gradient is turned on, creating a linear change in magnetic field strength from head to toe. Then an RF pulse with a specific frequency (corresponding to the desired slice location) is applied. Only protons in that slice precess at that exact frequency and absorb the RF energy — the slice is "selected."

Phase Encoding (Y-Gradient)

After slice selection, all spins in the slice are in phase and precessing at the same frequency (within the slice). To differentiate signals along the Y direction, a phase encoding gradient is briefly turned on. This gradient causes spins at one end of the Y-axis to speed up (higher field) and spins at the other end to slow down (lower field). When the gradient is switched off, all spins return to the base frequency, but they have different phases depending on their Y position. Each row of the image matrix will have a unique phase shift.

Frequency Encoding (X-Gradient)

After phase encoding, a frequency encoding gradient (also called the readout gradient) is applied along the X direction while the signal is recorded. This gradient causes spins at different X positions to precess at different frequencies — the signal contains a range of frequencies, each corresponding to a specific X location.

Signal Collection and Image Reconstruction

The receiver coil picks up the combined signal from the entire slice. This signal is a mixture of many frequencies and phases. The MRI system uses the known phase and frequency information to perform a Fourier transform, separating the signal into its constituent frequencies. Each frequency corresponds to a specific X position, and each phase corresponds to a specific Y position. The signal amplitude at each location is assigned a grayscale value:

  • White: strong signal

  • Black: no signal

The process is repeated for each phase encoding step (each Y row) until the entire matrix is filled. A standard matrix is 256×256256 \times 256256×256 or 512×512512 \times 512512×512, giving exquisite anatomical detail.

In essence: The three gradients allow the MRI scanner to encode the spatial origin of each NMR signal — slice selection (Z), phase encoding (Y), and frequency encoding (X) — making it possible to construct a 3D image of the body.