Electrode–Electrolyte Interface

Electrode–Electrolyte Interfaces and Electrochemical Cells

The electrical double layer

When a metal electrode is immersed in a water-based electrolyte, ions in the solution interact with the metal surface. Some ions may adsorb close to the electrode. This surface-adjacent layer of charge is called the inner Helmholtz plane (IHP). Ions of opposite charge are attracted toward it, producing a separation of charge across the electrode–solution boundary known as the electrical double layer.

Ions in water are generally solvated: water molecules surround them because water molecules are dipolar. Solvation affects how ions approach and interact with the electrode, although it is often omitted from simplified diagrams.

Ion charge density

An ion’s charge density depends on its charge relative to its effective volume. For example, a sodium atom loses an electron to form Na+\mathrm{Na^+}Na+. The resulting cation has a relatively small volume and a single positive charge, giving it comparatively high charge density. Chloride, Cl−\mathrm{Cl^-}Cl−, generally has lower charge density. As a broad tendency, cations often have higher charge density than anions, although there are exceptions.

Models of charge distribution

Different models describe how charge is arranged near an electrode:

  • Helmholtz model: Oppositely charged layers are treated as being separated by a short, fixed distance near the surface. In a simplified picture, the charge separation occurs over roughly 1 nm1\,\mathrm{nm}1nm.

  • Gouy–Chapman model: Charge is distributed through a more extended region of solution, potentially over a distance of about 10 nm10\,\mathrm{nm}10nm, rather than being confined to a thin layer.

  • Stern model: Combines features of both descriptions. Some charge is arranged close to the electrode in Helmholtz-type layers, while some is distributed farther into solution in a diffuse Gouy–Chapman layer.

The appropriate description depends partly on electrolyte concentration. At concentrations around 1 mol L−11\,\mathrm{mol\,L^{-1}}1molL−1, charge separation is largely concentrated near the electrode, so a Helmholtz-type description is often adequate. At much lower concentrations, such as approximately 1 mmol L−11\,\mathrm{mmol\,L^{-1}}1mmolL−1, the diffuse distribution described by the Gouy–Chapman model may be more relevant.

Double-layer capacitance and electric field

The double layer can be treated approximately as a capacitor, called the double-layer capacitor, with capacitance denoted CdlC_{\mathrm{dl}}Cdl​. A typical capacitance per unit area is about 101010–30 μF cm−230\,\mu\mathrm{F\,cm^{-2}}30μFcm−2.

Because charge is separated over a very short distance, the electric field can be extremely strong. A potential difference of 1 V1\,\mathrm{V}1V across a distance of 1 nm1\,\mathrm{nm}1nm corresponds approximately to

E=1 V1 nm≈109 V m−1.E = \frac{1\,\mathrm{V}}{1\,\mathrm{nm}} \approx 10^9\,\mathrm{V\,m^{-1}}.E=1nm1V​≈109Vm−1.

Thus, substantial potential changes can occur across a nanometre-scale region at the interface.


Interfacial processes and equivalent circuits

The double-layer capacitor describes charge separation when no electron-transfer reaction occurs across the interface. At a sufficiently suitable potential, however, an electrochemical reaction can take place. A species in the electrolyte may gain or lose an electron, material may dissolve from the electrode, or material from the electrolyte may deposit onto it.

An electrochemical reaction involves both a chemical change and electron transfer. Such a reaction produces current and is called a faradaic process. Its electrical behavior can be represented by a faradaic impedance, denoted ZFZ_{\mathrm{F}}ZF​. In a simplified interfacial circuit, the double-layer capacitance and faradaic impedance are in parallel:

  • CdlC_{\mathrm{dl}}Cdl​ represents charging of the electrical double layer.

  • ZFZ_{\mathrm{F}}ZF​ represents current associated with electrochemical reactions.

If no appreciable reaction occurs, the interface may be represented mainly by CdlC_{\mathrm{dl}}Cdl​. If a reaction does occur, both capacitive and faradaic current may contribute.

A two-electrode cell model

A basic electrochemical cell requires at least two electrodes, with electrolyte between them. Each electrode–electrolyte interface can be represented by its own double-layer capacitance and faradaic impedance. The electrolyte also resists current flow; this contribution is represented by the solution resistance, RsolR_{\mathrm{sol}}Rsol​.

A simple cell model therefore includes:

  • The first interface, represented by Cdl,1C_{\mathrm{dl,1}}Cdl,1​ and, when relevant, ZF,1Z_{\mathrm{F,1}}ZF,1​.

  • The electrolyte resistance, RsolR_{\mathrm{sol}}Rsol​.

  • The second interface, represented by Cdl,2C_{\mathrm{dl,2}}Cdl,2​ and, when relevant, ZF,2Z_{\mathrm{F,2}}ZF,2​.

This model emphasizes that measured cell behavior can involve both interfaces and the solution between them.


Why the potential of a single electrode cannot be measured directly

Immersing a metal in an electrolyte creates a potential difference across its interface with the solution. However, the potential of that single interface cannot be measured independently with an ordinary voltage measurement.

A voltmeter needs two connections. If one connection is attached to the electrode and the other is placed in the electrolyte, the second connection also forms an electrode–electrolyte interface and therefore develops its own potential difference. The measured voltage is consequently a difference involving both interfaces, not the potential of just one.

If the potential drop across the interface of interest is xxx and the potential associated with the measuring connection is yyy, the instrument measures a quantity such as x−yx-yx−y. Since yyy is not generally zero or negligible, the absolute value of xxx cannot be obtained from that measurement alone.

A further difficulty occurs in a two-electrode cell. The applied voltage is distributed across both electrode interfaces and the electrolyte. Even if the total applied voltage is known, the individual potential drops are not automatically known. For example, changing the applied cell voltage does not guarantee that the additional voltage will be shared equally between the two interfaces. The division depends on the behavior of the cell.

Consequently, a two-electrode measurement may show how the current changes as the total applied voltage changes, but it does not by itself reveal how much the potential changed at either individual interface. This makes it difficult to interpret reactions at a particular electrode.


Reference electrodes and working-electrode potential

A reference electrode addresses this measurement problem. It is designed to maintain a stable, reproducible potential difference at its own interface. Its absolute interfacial potential still is not measured directly; rather, its stable behavior provides a dependable comparison point.

The electrode being studied is called the working electrode. When a reference electrode is used with a working electrode, changes in the measured potential can be assigned to changes at the working electrode, provided that the reference potential remains stable. In particular, changing the applied potential changes the working-electrode potential relative to the reference in a known way, even though the absolute potential of either single interface is not directly known.

Electrode potentials must therefore be reported relative to a specified reference electrode. Common examples include:

  • Standard hydrogen electrode (SHE)

  • Standard calomel electrode (SCE)

  • Silver/silver chloride electrode (Ag/AgCl), used with specified potassium chloride concentrations

Results reported against different reference electrodes can be translated between scales using a constant potential offset, provided the reference electrodes and conditions are properly specified.

Luggin capillary

The solution between the working electrode and the reference electrode has resistance. To reduce the effect of this resistance on the measured potential, the reference electrode can be positioned close to the working electrode without otherwise disturbing the system.

A Luggin capillary helps place the reference-electrode tip near the working electrode. It can be filled with the electrolyte and positioned so that its tip is close to the working surface. This reduces the solution-resistance contribution between the working electrode and the reference point.


The three-electrode system

A practical three-electrode arrangement contains:

  1. Working electrode: The electrode at which the reaction or interface of interest is studied.

  2. Reference electrode: Provides a stable comparison potential and is used to measure the working-electrode potential.

  3. Counter electrode, also called the auxiliary electrode: Carries the current required by the experiment.

The reference electrode should carry little or no significant current. If substantial current passes through it, its interfacial conditions can change and it may no longer maintain a constant potential. The counter electrode supplies or accepts the current instead, so the reference electrode can remain a stable potential reference.

In the arrangement described, the instrument measures and controls the potential between the working and reference electrodes, while current flows between the working and counter electrodes. The instrument adjusts the counter-electrode potential as needed to support the required current.

Counter-electrode area and material

The counter electrode is commonly made with a large surface area relative to the working electrode. For a given current, a larger counter-electrode area lowers the current density at that electrode. This reduces the potential demands placed on the counter electrode and on the instrument. If the counter electrode is too small, it may need to operate at a much higher current density, potentially creating unwanted effects or exceeding the instrument’s operating limits.

Platinum mesh is a common choice because platinum is relatively inert for many applications, while the mesh provides a large surface area in a compact geometric space.

Applied voltage and compliance voltage

Electrochemical instruments, such as potentiostats, are specified using two distinct voltage ranges:

  • Applied voltage range: The range of potential that can be applied between the working electrode and the reference electrode.

  • Compliance voltage range: The voltage the instrument can provide between the working and counter electrodes to drive the required current.

The instrument must respond rapidly and stably as the counter-electrode potential is adjusted. It uses closed-loop control to manage this response and must measure currents that may be as small as microamperes, nanoamperes, or even picoamperes. These control and measurement requirements make a three-electrode potentiostat more complex than a basic direct-current power supply.

A two-electrode arrangement can be sufficient for ordinary electrical-circuit analysis or certain special cases. For many electrochemical experiments, however, a three-electrode system is preferable because it allows the working-electrode potential to be controlled relative to a stable reference while another electrode carries the current.


Electrochemical response and the water-reaction window

For a simple resistor, changing the applied voltage produces a current change described approximately by Ohm’s law:

VI=R.\frac{V}{I}=R.IV​=R.

In many ordinary electrical systems, resistance is approximately constant under fixed conditions, although it can vary with factors such as temperature. Electrochemical systems can behave very differently: over some voltage ranges, a substantial change in applied potential may produce little current, while at other potentials the current can rise sharply because reactions begin to occur.

Water splitting limits the usable potential range

In an aqueous electrolyte, applying sufficiently extreme potentials can cause water to decompose:

  • At one electrode, oxygen can evolve.

  • At the other electrode, hydrogen can evolve.

Reversing the polarity reverses which electrode is driven toward each reaction. If an experiment is intended to study a different reaction, it is generally important to operate within a potential range where water splitting does not dominate. The range of potentials available for other processes is therefore limited by the onset and rate of hydrogen and oxygen evolution.

Thermodynamics, kinetics, and electrode material

Thermodynamics indicates whether a reaction is favored beyond a given potential. It does not, by itself, establish how quickly the reaction proceeds. The reaction rate depends on kinetics, including how readily the electrode surface catalyzes the process.

For hydrogen evolution, platinum is a good catalyst, so hydrogen can form readily. Zinc, lead, and mercury are comparatively poor catalysts for this reaction. With these less active surfaces, hydrogen production can be slow even when the thermodynamic conditions favor it. This kinetic delay can allow experiments over a wider potential range before hydrogen evolution becomes significant.

The practical window can also be extended by changing the solvent. Non-aqueous electrolytes can permit a different usable range, but every solvent has a breakdown limit. Applying potentials beyond that limit causes solvent decomposition rather than providing an indefinitely wider operating range.


Supporting electrolyte and solution resistance

A supporting electrolyte is a salt or acid added primarily to increase the solution’s conductivity. Ideally, it does not participate in the reaction being studied and does not otherwise change the system in a relevant way. Sodium perchlorate is an example mentioned for this purpose.

Adding sufficient supporting electrolyte reduces the solution resistance, RsolR_{\mathrm{sol}}Rsol​. With lower solution resistance, less of the applied potential is lost across the bulk solution, and a larger share of the potential change occurs across the working-electrode interface. This makes the working-electrode potential more effectively controlled and interpreted.

When conductivity is low—for example, in water containing only a small amount of salt—the resistance of the solution can be substantial. Current through the cell then produces a significant potential drop in the electrolyte. As a result, the total applied voltage is not confined to the electrode interface: some of it may be lost between the electrodes in the solution. The working-electrode interface may therefore experience a smaller or different potential change than the applied voltage alone suggests.

With abundant supporting electrolyte, the solution resistance can become small enough to neglect approximately in a simplified circuit. The working electrode can then be modeled mainly by its double-layer capacitance and faradaic impedance, while the reference electrode contributes little impedance. Supporting electrolyte is thus used not to create the target reaction, but to improve conductivity and reduce unwanted potential losses in the solution.