Atomic Emission Spectrometry Sources

Transcript

The ideal source for atomic emission   spectrometry should have the following properties.  1. Temperature should be high enough. That means,   the plasma should be very stable. The temperature  should be high enough for all the elements you   want to analyze must be excited to the next  higher energy level. That’s what it means. That is a universal applicable principle. 2. The background intensity should be as low   as possible for high detectability. 3. The temperature should be uniform   and the optical density should be low  for linear calibration curves.   4. The rate at which the sample is  acceptable should be reasonable. That   means, the representative sub sample it  should be taken in an acceptable time.   5. The sample should be acceptable  as solid, liquid or gaseous forms.

6. The introduction of the sample should  not drastically alter the internal energy   of the source. That means, the  plasma should be very stable.   7. All the sample material should be  converted into free atoms within the source.   8. The source would operate on commonly available,  reasonably priced gases that is argon only,   as far as our experience is concerned. The ICP comes close to fulfilling these specifications for an ideal source. Because, 1. The temperature is high enough and majority   of the elements can be determined 2. Background is really low. Because   optically it is almost transparent. 3. The temperature is fairly uniform   and the tail flame is optically thin. That  means, not much emission coming from the   tail flame. The background would be very less. 4. Sample is acceptable in all the forms.

5. The sample input through the peristaltic  pump is acceptable at a reasonable rate.   6. The introduction of the sample  does not seriously affect the energy   7. For gaseous and liquid samples the  material is converted into atoms   8. The plasma operates an argon which is  freely available at a reasonable price.   It has some advantages, 1. Rapid and simultaneous   multi element analysis. 2. It is an excellent source of atomic

emission that it is optically thin. 3. Calibration curves are linear   over 4 to 6 orders of magnitude. 4. Detection limits are very low: 1 to 100 ng   per g or μg per l (parts per billion) 5. It is suitable for analyses at all   concentrations from trace to major levels 6. Accuracy and precision are very good   7. In principle, ICP-AES can be used for  the determination of all elements except   argon. In practice, approximately  70 element can be determined using

ICP-AES with 1 single source that is plasma. 8. Elements that are difficult to determine by   AAS can be determined easily for example, boron  carbon, cerium, lanthanum etc. These are rare   earth elements. Then actinium elements  and refractory elements like titanium,   tantalum, vanadium, zirconium, etcetera. Now I want to talk to you about analytical   aspects. Because, any chemical analysis  must follow certain fundamental principles

of chemical analysis. I want to define  certain terms for your easy reference.

The reference what I am giving you is  the International union of Pure and   Applied chemistry (IUPAC). International  organization for standardization that   is ISO6955, this defines the Analytical  Spectroscopic methods for Flame emission,   Atomic Absorption and Atomic Fluorescence. Solid samples can be directly analyzed by AES,   but the majority of the samples are  presented as liquids as solutions.

Sample solution is a solution, suitably  made up from a test portion of the sample   submitted for analysis. That means, it is  the solution to be aspirated into the ICP. Stock solution is a solution of known  concentration from which the dilutions are   made to prepare the calibration standards. Reference solution are series of known   concentrations of the stock solutions. A series  of known concentration of the analyte element   in a solvent is known as reference solution.  It may contain the matrix elements also.

Blank solution is a 0 member compensation  solution containing all other elements that   is matrix matching in the same solution, but  it does not contain the analyte. Blank solution   should be having 0 concentrations of the analyte. Analytical curve is the calibration curve or a   graphical plot of the measured absorbance or  emission to the concentration or mass of the   analyte sample. This is the curve  what you need for interpretation

of the results during chemical analysis. Characteristic concentration or characteristic   mass. This refers to the concentration  of the analyte element corresponding to   a net absorption or emission of 1 percent or an  absorbance of 0.044, when integrated absorbance   or peak area is used. We can take almost  the same for emission ICP-AES also.

Sensitivity is the slope of the analytical curve  that is termed as sensitivity. It is delta a by   delta c or delta a by delta m. delta c is the  concentration, delta a is the absorbance or   emission. Change in the emission or absorption  with respect to concentration or mass is know   as sensitivity. For good sensitivity, change  should be large for small concentration change.

Accuracy relates to the closeness of the agreement  between the true value of an element in a sample   and the mean value. Usually true value is not  determinable. But mean value is determinable.   It can be calculated by the difference  between the true value and the measured   value. There will be some slight difference.  That difference should be as low as possible   compared to the true value and mean value. Precision relates to the closeness of the

agreement between the results obtained by applying  the analytical procedure repeatedly. Suppose you   do three different analysis using the same set  of conditions. If the results match, then we say   results are very precise. It does not mean results  are correct. Because, the same error could have   been carried out in all the three samples.  So, a precision does not guarantee accuracy.   But precision means the system is perfect. It is  determined by multiplying the standard deviation σ with 2.83 for 30 or more measurements  or t.σ, where t is the student’s factor.

In Student factor “Student” is  the name of the scientist. Standard deviation is given by square root  of x y minus x bar whole square divided by   n minus 1. Where x bar is the mean value of  all measurements and xi is the individual   measurement. n is the number of measurements.  In standard deviation what we mean is,   the actual analytical value is spread over the  mean value plus or minus some number. It is a very   important concept in all analytical measurements. Relative standard deviation can be calculated by

dividing standard deviation by the mean value. Determination limit is the lowest concentration   that can be determined with the prescribed  precision for practical procedure. That means,   not detected it is determined. Quantitative  determination is involved here.   . Detection limit   is delta c by delta a into k dot sigma or delta m  by delta a into k dot sigma. The factor k is 2 or   3. Sigma is the absolute standard deviation. The  detection limit can be measured either using peak

area or peak height. That is the emission line  spectrum. Both are functions of electrical noise,   optical noise, sensitivity, metastable energy  status of the analyte element and the chemical   environment. That is matrix matching. The detection limit can vary for different   systems. It should be evident to you that, the  reliability of a measurement depends upon all   these factors. Detection limit, standard  deviation, relative standard deviation,

slope, sensitivity etc., we should be  able to determine all these things.   I have a figure here for you. That  makes a difference here. For example,   this is the mean background signal. Here is  the intensity signal, number of counts. This   is your sample, the signal goes like this.  Three times about this value of the mean   background signal is the limit of detection.  Intensity signal limit of identification.

So, we can say if you get a signal around this  level, element maybe there. But to make it sure   that, the element of analysis is there. We  have to go to a higher level that is known as   limit of identification. If you get a reading  like this. It means the element is definitely   there. How much is there is gives you the limit  of determination. Above this if you get a reading   anywhere around this region, the determination  is possible. So, you have to understand that,

this is not done according to scale. Because,  I will have to define all the k sigma,   2.3 student factor t and all those things. The determination limit that can be determined   with the prescribed precision  for the analytical procedure. Calibration technique. AAS is the relative  technique and not an absolute technique.

This means that quantitative result can  be obtained by comparison with reference   solutions or reference materials. Therefore  reference measurements must always be made.   Different calibration techniques are available  to meet the analytical requirement and internal   standards need to be employed in AES. The analytical curve technique,   bracketing technique and analyte addition  technique all these things can be employed to determine the concentrations of the samples. The wide dynamic range of ICPS is certainly   impressive. Calibration graphs are essentially  linear below 10 ppm even upto 1000 ppm. Stock   solution itself can be measured if the sensitivity  is low. As we noted previously, this is not only   desirable attribute of the technique, but it  is essential for us, if the full potential of   simultaneous analysis is to be exploited. Sometimes what happens is, there we need to

extract and enrich the sample and separation.  Insufficient sensitivity is 1 of the reasons   faced with such a problem. A more sensitive  method needs to be employed. In trace and   ultra-trace analysis every sample pretreatment  is a potential source of contamination,   we have to be very careful. If separation and  enrichments are time consuming, they do demand   a degree of knowledge and skill. This is where  the chemical ingenuity of the person performing

the chemical analysis comes into play. In trace and ultra-trace analysis every sample   pretreatment should be a source of contamination.  As far as possible we try to avoid. That is the   reason we try to introduce the sample as it is. Among the techniques used for separation and   enrichment of the analyte solvent extraction,  activated charcoal method, the electrolytic   decomposition etc., are prominent. Chelating agents such as EDTA,

ammonium pyrrolidine, dithiocarbamate, NTA  etc., are used to complex many metals over   a wide pH range followed by extraction in  to MBK, (methyl ethyl ketone) or any other   solvent known to facilitate trace analysis. Preconcentration with activated charcoal is very   well known example of enriching the analyte.  Quite often the samples need to be dissolved in   acids under pressurized atmosphere. Any sample  treatment for that matter can be a potential

source of error, if we do not take the proper  precautions. Ultra-pure water and acids need to   be employed for the analysis as well as in parts  per million or parts per trillion level. It is for   AAS as well as ICP. Among the techniques used for  separation and enrichment of the analyte sample   solvent extraction etc., are prominent. Specific applications of atomic emission   spectrometry. Here is a list of about ten areas.  The determination of calcium, magnesium, sodium,

potassium zinc in body fluids and tissues.  Tissue samples must be wet and dry ashed   before analysis. Nowadays aluminum is also an  element of focus in the brain tissues owing to   Alzheimer’s disease. In biological applications  body fluids, arsenic mercury, cadmium, selenium,   thallium, etc., are of importance. In food and drinks, we have animal products   as such as ghee, butter, cheese and metallic  elements are present in many of the foodstuffs.

Functional groups attached to tissues they need  to be digested with nitric acid. Almost all dairy   products must be wet ashed before analysis. Lead  a very common contaminant in foodstuffs. Great   emphasis is being laid on the their analysis of  lead. Tin is also added to plastic foils.   Soil, fertilizer and plants. Sodium,  potassium, calcium, magnesium, copper,   manganese, zinc, boron, arsenic, cadmium,  lead, all these are of importance in the

soil analysis for using ICP AES. We determine almost all elements in   water. Sometimes you use APDC and MIBK for  concentrating the elements of interest. Many   other extraction systems have also been  perfected over the years. The details are   available in the standard textbooks. For sediment  digestion, we need HF and aquaregia are difficult   to dissolve the samples metals. In the environment, it covers a whole range of factors that influence our lives. It  is almost impossible to pick out any specific   area in isolation. Since all factors  influence each other mutually.

Here is a list of possible interactions in  the general word. You can see that many of   these are all equilibrium reactions. Air, humans,  elements, commodities, Drinking water, foodstuff,   plants, animals, surface water, and subterranean  water, waste water, garbage and soil.

A majority of the particulate matter in  the air originates from natural sources,   dust and aerosols of 2.5 micron size, enter our  lungs, they deposit in our lungs. Stock gases   are known to carry particulate matter even up to  400 kilometer. Suppose, you have a factory that is   spewing out the smoke, the smoke components can  be carried out up to 400 kilometers away also.

Air pollutants are normally found everywhere.  You cannot escape from the polluted air. Most   of the coal gas fired pollutants get enriched  in the flue dust of treated gases cadmium and   lead are collected by filtration techniques over  4 hour time sometimes 6 hours sometimes 24 hours   depending upon the requirement. Toxic heavy metals  including mercury are passed on to soils surface,   water etc. Sewage sludge and electronic  commodities require digestion anyway.

ICP was developed mainly for the analysis of  rocks and minerals in ores. Calcium, magnesium,   potassium, iron, copper, zinc, gold, aluminum,  platinum, many of these elements have been used.   ICP-AES and hydride techniques have substantially  contributed to the determination of trace   elements. Digestion with HF and sulfuric acid is  the most preferred technique for minerals.   For volatile elements such as mercury, cadmium  etc., what we need is, pressurization using

power bomb. The use of common fusion procedure  using sodium carbonate, sodium bromide followed   by dissolution of the melt in acid the molybdate  solution is well known. Gold is usually extracted   and enriched before determination because its  concentration is too low for direct analysis   Metallurgy and plating. Aluminum, vanadium,  titanium, chromium, niobium and all these   elements are usually determined. If you want to  determine plating solutions HCI/HNO3 treatment is

a fairly suitable technique. ICP-AES is suitable  for trace elements, bismuth, lead, selenium,   tellurium, thallium in complex nickel alloys. Bell  has investigated the determination of cadmium,   chromium, copper, iron, magnesium, in aluminum  alloys. They dissolve the alloy simply in 1:1 HCl   with the addition of few drops of H2O2. Oil, coal and petrochemistry. It is useful for   the elements of trace analysis in oil and  petroleum products. Iron and copper are the

additives in the lube oil which are monitored  quite regularly in ppm quantities by flame flame   or ICP AES. Lead in petrol is the favorite  topic of all environmental analysts   Plastic paper and textiles. Iron, aluminum,  manganese, titanium, silicon, antimony, led, tin,   zinc, and many of these things are determined  in plastics nowadays. The polypropylene,   nylon 66, polystyrene, PVC, cyclohexane,  formic acid, etc all these polymers are

dissolved in MIBK and aspirated directly. Insoluble polymers are need to be digested   and bring them in solution. Packing materials  may contain organotin compounds as stabilizers.   Metal plastic sample is extracted with  3% acetic acid. Heptane ethanol etc.,   extracts are determined directly in ICP AES.  That is why I keep on telling you that matrix   matching is a very important concept in ICP  AES. Otherwise there is no way we can determine

such elements in such low concentration  zones. Langmyer has determined cadmium,   copper, lead and manganese in paper, pulp,  cellulose directly. That is why their work   in such matrices is very well known. Pharmaceuticals, miscellaneous products,   industrial products, quite a few elements are of  importance. Sodium, potassium, calcium, arsenic,   cobalt, manganese, silver, etc. The samples need  to be dissolved or leached in concentrated HCl

and aspirated directly. Cobalt in vitamin B12  and proteins can be determined by the flame   AAS as well as ICP AES. Barium can enter from the  glows or rubber components in addition to direct   ingestion. Diet supplements are routinely  checked for arsenic and selenium. Uranium in yellow cake is usually determined by  ICP-AES. Many other elements like lead, magnesium,   iron, potassium, cadmium, copper are routinely  determined in rubber seals, radioactive elements, ICP-AES is the panacea. The available literature  up to date is the proof of its vivacity only.

Here is a typical flow chart showing the  methodology for a typical ICP. How do you   go about actually doing the analysis. First  prepare the samples and standards. Then find   out is there a ready-made method for measuring  the wave length and concentration. If it is   there you have to tell the computer to recall  the method. Put all the parameters in place. Aspirate your calibrated samples using standard  solutions. Then run the samples. Generate the   report. This is how the normal scheme  of chemical analysis using ICP AES run.

Suppose, there is no method. Then what do you  do? We will have to select elements and then   collect the representative spectra. Select  integration time, data acquisition mode and   standard concentrations etc. Then select the  background, store the method and then go back to   developing a method using this technique. This is how Sequential ICP AES looks. Nowadays   it can be put on a bench top. It is a  very simple instrument to operate. On

the right side you can see the plasma.  Inside the plasma there is a toroidal   space. The sample comes from the bottom. I would like to thank you for your attention   during all these lectures. I hope, I have given a  sort of preliminary introduction to the technique   of ICP AES. If you need any additional information  or any help you can always contact me.

I wish you all the best. Thank you very much.