Atomic Emission Spectrometry Sources
- 113Unfamiliar
- 0Learning
- 0Familiar
- 0Mastered
113 questions
- 1For an ideal atomic emission source, the plasma must be hot enough to excite every analyte element to the_____.
Hint
Think about the energy transition required for emission.
Answer
next higher energy level
Explanation
A sufficiently high temperature is needed to excite all elements being analyzed to the next higher energy level.
- 2A stable plasma is important because the source temperature should remain_____.
Hint
Consider whether temperature fluctuations would support reliable excitation.
Answer
stable
Explanation
The ideal source should have a very stable plasma so its temperature and excitation conditions remain consistent.
- 3To improve detectability, the ideal source should have background intensity that is as_____as possible.
Hint
A weaker background makes analyte signals easier to distinguish.
Answer
low
Explanation
Low background intensity improves the detectability of analyte emission.
- 4A uniform temperature and low optical density help produce_____calibration curves.
Hint
These source properties support a proportional signal response.
Answer
linear
Explanation
The source specifications associate uniform temperature and low optical density with linear calibration curves.
- 5The ideal source should accept representative subsamples at a_____rate.
Hint
The sample throughput should be practical rather than excessively slow.
Answer
reasonable
Explanation
The source should allow a representative subsample to be taken and introduced in an acceptable amount of time.
- 6An ideal source should be able to accept samples in solid, liquid, or_____form.
Hint
This is the third common state of matter listed with solids and liquids.
Answer
gaseous
Explanation
The ideal source should accommodate samples in solid, liquid, or gaseous forms.
- 7Introducing a sample should not drastically alter the source's_____.
Hint
The relevant property describes the energy state within the source.
Answer
internal energy
Explanation
An ideal source remains stable when a sample is introduced, without a drastic change to its internal energy.
- 8Ideally, all sample material entering the source should be converted into_____.
Hint
The desired products are uncombined atoms that can produce atomic emission.
Answer
free atoms
Explanation
The ideal source converts all of the sample material into free atoms.
- 9The ideal source should use commonly available, reasonably priced_____.
Hint
These are used to operate the plasma source.
Answer
gases
Explanation
Commonly available and reasonably priced operating gases are a desired source property.
- 10In the described ICP, the commonly used plasma gas is_____.
Hint
It is a noble gas used to sustain the plasma.
Answer
argon
Explanation
The source describes the ICP as operating on argon, which is reasonably priced and readily available.
- 11The ICP comes close to fulfilling the specifications for an_____atomic emission source.
Hint
This is the standard against which the source properties are compared.
Answer
ideal
Explanation
The ICP is presented as coming close to the desired properties of an ideal source.
- 12The ICP is optically almost transparent, which helps keep its_____low.
Hint
A low value of this signal makes analyte emission easier to detect.
Answer
background
Explanation
The ICP's near transparency is given as a reason for its very low background.
- 13The ICP tail flame is described as optically thin, so it produces little_____.
Hint
The low signal from the tail flame reduces the unwanted baseline.
Answer
emission
Explanation
The optically thin tail flame contributes little emission, keeping background low.
- 14The sample is introduced into the ICP at a reasonable rate using a_____pump.
Hint
This pump type moves liquid by compressing and releasing flexible tubing.
Answer
peristaltic
Explanation
The described ICP sample input uses a peristaltic pump to deliver the sample at a reasonable rate.
- 15ICP-AES allows rapid and_____analysis of multiple elements.
Hint
Several elements can be measured during the same analysis.
Answer
simultaneous
Explanation
Rapid simultaneous multi-element analysis is listed as an advantage of ICP-AES.
- 16The ICP is an excellent atomic emission source in part because it is optically_____.
Hint
This property limits self-absorption and background effects.
Answer
thin
Explanation
The source identifies the ICP as an excellent emission source because it is optically thin.
- 17The stated ICP-AES detection limits are about one to one hundred nanograms per gram or micrograms per liter, corresponding to_____.
Hint
This concentration scale is abbreviated ppb.
Answer
parts per billion
Explanation
The described detection limits are in the parts-per-billion range.
- 18ICP-AES is suitable for measuring concentrations from trace levels through_____levels.
Hint
This is the high-concentration end of the stated range.
Answer
major
Explanation
The technique is described as suitable across concentrations from trace to major levels.
- 19The performance of ICP-AES is described as having very good accuracy and_____.
Hint
This property concerns the agreement among repeated measurements.
Answer
precision
Explanation
Very good accuracy and precision are both listed among the advantages of ICP-AES.
- 20In principle, ICP-AES can determine all elements except_____.
Hint
This is also the gas used to operate the plasma.
Answer
argon
Explanation
The source states that ICP-AES can in principle be used for all elements except the plasma gas argon.
- 21In practice, about_____elements can be determined using a single ICP-AES plasma source.
Hint
The source gives an approximate count rather than a complete list.
Answer
70
Explanation
The described practical capability is approximately seventy elements using one plasma source.
- 22Elements that can be difficult to determine by AAS but are readily determined by ICP-AES include boron, carbon, and_____.
Hint
Cerium and lanthanum are examples of this group.
Answer
rare earth elements
Explanation
The lecture gives boron and carbon as examples and also mentions rare earth elements such as cerium and lanthanum.
- 23Titanium, tantalum, vanadium, and zirconium are examples of_____elements that ICP-AES can determine.
Hint
These elements are characterized by resistance to heat and difficult processing.
Answer
refractory
Explanation
The source lists titanium, tantalum, vanadium, and zirconium as refractory elements that can be determined by ICP-AES.
- 24ICP calibration graphs are essentially linear below 10 ppm and can extend up to about_____ppm.
Hint
This upper value is stated as one thousand parts per million.
Answer
1000
Explanation
The described wide dynamic range includes essentially linear calibration below 10 ppm and up to about 1000 ppm.
- 25When sensitivity is low, even the_____solution can sometimes be measured directly.
Hint
It is the concentrated solution from which dilutions are prepared.
Answer
stock
Explanation
The wide dynamic range may allow the stock solution itself to be measured if the sensitivity is low.
- 26A solution prepared suitably from the test portion submitted for analysis is called the_____solution.
Hint
It is the prepared solution that is aspirated into the ICP.
Answer
sample
Explanation
The sample solution is made from the test portion and is the solution presented for analysis.
- 27A solution of known concentration used to prepare dilutions for calibration standards is a_____solution.
Hint
It serves as the starting solution for making standards.
Answer
stock
Explanation
A stock solution has a known concentration and is used to prepare diluted calibration standards.
- 28A series of solutions with known analyte concentrations in a solvent is called_____solutions.
Hint
These known solutions provide comparison values for analysis.
Answer
reference
Explanation
Reference solutions are a series of known analyte concentrations in a solvent.
- 29Reference solutions may contain the analyte and also the sample's_____elements.
Hint
These are the other components surrounding the analyte in the sample.
Answer
matrix
Explanation
Reference solutions can include matrix elements in addition to the analyte.
- 30A blank solution contains the matrix components but has zero concentration of the_____.
Hint
This is the element whose concentration is being measured.
Answer
analyte
Explanation
The blank is a matrix-matched solution that does not contain the analyte.
- 31A graphical plot of measured absorbance or emission against analyte concentration or mass is the_____.
Hint
It is also called the calibration curve.
Answer
analytical curve
Explanation
The analytical curve relates measured absorbance or emission to the concentration or mass of the analyte.
- 32The concentration or mass corresponding to a net signal of one percent, or absorbance of 0.044, is called the_____.
Hint
This term identifies a signal response at a specified small level.
Answer
characteristic concentration or characteristic mass
Explanation
Characteristic concentration or characteristic mass refers to the analyte level corresponding to the stated net signal or absorbance.
- 33When integrated absorbance or peak area is used, the absorbance specified for characteristic concentration is_____.
Hint
It is the numerical absorbance value stated alongside a net one-percent signal.
Answer
0.044
Explanation
The source gives 0.044 absorbance as the characteristic value when integrated absorbance or peak area is used.
- 34The slope of the analytical curve is called its_____.
Hint
It describes how much the signal changes as analyte amount changes.
Answer
sensitivity
Explanation
Sensitivity is defined as the slope of the analytical curve.
- 35Good sensitivity means a large change in signal for a_____change in concentration.
Hint
A method is more responsive when little analyte change produces a strong signal change.
Answer
small
Explanation
Good sensitivity requires the signal to change substantially when concentration changes only slightly.
- 36Accuracy describes how closely a measured mean agrees with the_____value.
Hint
This is the value the analytical result aims to estimate.
Answer
true
Explanation
Accuracy concerns the closeness of agreement between the true analyte value and the mean measured value.
- 37Precision describes the agreement among results obtained by repeating the analytical procedure under the same_____.
Hint
Keep the procedure and measurement setup consistent in the repeated analyses.
Answer
conditions
Explanation
Precision is the closeness of agreement among repeated results obtained using the same analytical procedure.
- 38Repeated results can be precise without being accurate if they all contain the same_____.
Hint
A consistent systematic problem can make results agree with each other but miss the true value.
Answer
error
Explanation
Precision does not guarantee accuracy because the same error may be present in every repeated measurement.
- 39The standard deviation describes the spread of measurements around their_____.
Hint
It is the central average of the measurement set.
Answer
mean value
Explanation
Standard deviation characterizes how measurements are distributed around their mean.
- 40Relative standard deviation is calculated by dividing standard deviation by the_____.
Hint
It scales the spread relative to the central measurement.
Answer
mean value
Explanation
Relative standard deviation is the standard deviation divided by the mean value.
- 41The lowest concentration measurable with the prescribed precision in a practical procedure is the_____.
Hint
At this level, the analyte is quantitatively determined rather than merely noticed.
Answer
determination limit
Explanation
The determination limit is the lowest concentration that can be measured with the prescribed precision.
- 42The detection limit is calculated from signal sensitivity and the absolute standard deviation, multiplied by a factor_____.
Hint
The source assigns this factor a value of two or three.
Answer
k
Explanation
The detection-limit expressions use a factor called k together with absolute standard deviation and signal sensitivity.
- 43In the detection-limit calculation, the factor k is usually_____or three.
Hint
The two possible values are small whole numbers.
Answer
2
Explanation
The stated value of k is two or three.
- 44The sigma used in the detection-limit calculation represents the_____standard deviation.
Hint
It is not the relative standard deviation.
Answer
absolute
Explanation
Sigma in the detection-limit expression is the absolute standard deviation.
- 45A detection limit may be measured using either peak area or peak_____.
Hint
This is the other common measure of an emission-line signal.
Answer
height
Explanation
Detection limits can be determined from either peak area or peak height.
- 46The detection limit depends on electrical noise, optical noise, sensitivity, analyte metastable energy status, and the chemical_____.
Hint
This includes the matrix surrounding the analyte.
Answer
environment
Explanation
These factors, including the analyte's chemical environment and matrix matching, affect the detection limit.
- 47Because detection limits vary among systems, measurement reliability depends on factors such as standard deviation, relative standard deviation, and_____.
Hint
This is the response of signal to a change in analyte concentration or mass.
Answer
sensitivity
Explanation
Reliability depends on several measurable characteristics, including detection limit, standard deviations, slope, and sensitivity.
- 48A signal around three times the mean background signal is used as an estimate of the limit of_____.
Hint
At this threshold, the element may be present but is not necessarily confirmed quantitatively.
Answer
detection
Explanation
The illustrated limit of detection is approximately three times the mean background signal.
- 49A signal higher than the detection threshold that confirms the element's presence is called the limit of_____.
Hint
This threshold provides stronger evidence that the analyte is present.
Answer
identification
Explanation
The limit of identification is a higher signal level used to make sure the element is present.
- 50The threshold at which the amount of an analyte can be measured quantitatively is the limit of_____.
Hint
This is the level at which determining how much analyte is present becomes possible.
Answer
determination
Explanation
Above the limit of determination, quantitative measurement is possible with the prescribed precision.
- 51For thirty or more measurements, the described precision estimate uses standard deviation multiplied by_____.
Hint
This is the numerical multiplier specified for large measurement sets.
Answer
2.83
Explanation
The source gives 2.83 times the standard deviation for thirty or more measurements.
- 52For other measurement counts, the described precision estimate can use t multiplied by the standard deviation, where t is the_____factor.
Hint
This factor is named after a scientist and depends on the statistical context.
Answer
Student's
Explanation
The source refers to t as the Student's factor.
- 53In a standard deviation calculation, the number of measurements is represented by_____.
Hint
It is the count of observations in the measurement set.
Answer
n
Explanation
The source identifies n as the number of measurements used to calculate standard deviation.
- 54The average of all measurements in a standard deviation calculation is the_____.
Hint
It is the central value around which individual results vary.
Answer
mean value
Explanation
The mean value is the average of all measurements and is used to describe their spread.
- 55AAS is described as a relative technique rather than an_____technique.
Hint
Its quantitative results are obtained by comparison rather than independently.
Answer
absolute
Explanation
AAS is a relative technique: quantitative results come from comparison with reference solutions or materials.
- 56Because AAS is a relative technique, quantitative analysis requires comparison with reference solutions or reference_____.
Hint
These provide known values for comparison with the sample.
Answer
materials
Explanation
Reference measurements must be made using reference solutions or reference materials.
- 57The analytical curve, bracketing, and analyte addition are all calibration_____.
Hint
They are different ways to establish or apply a concentration response.
Answer
techniques
Explanation
The source lists the analytical curve, bracketing, and analyte addition as techniques for determining sample concentrations.
- 58In AES, the use of_____standards is identified as a calibration need.
Hint
These standards are introduced to help account for variation during measurement.
Answer
internal
Explanation
The source states that internal standards need to be employed in AES.
- 59When ICP sensitivity is insufficient for trace analysis, a more_____method may be needed.
Hint
The alternative must respond more strongly to small analyte quantities.
Answer
sensitive
Explanation
Insufficient sensitivity can require a more sensitive analytical method or a sample enrichment step.
- 60In trace and ultra-trace analysis, every sample pretreatment is a potential source of_____.
Hint
Even small amounts introduced during handling can affect results.
Answer
contamination
Explanation
Sample pretreatment can introduce contamination, so it should be minimized and carefully controlled.
- 61One way to reduce contamination risk is to introduce the sample as it is and avoid unnecessary_____.
Hint
This refers to processing the sample before measurement.
Answer
pretreatment
Explanation
The source emphasizes avoiding sample pretreatment whenever possible in trace and ultra-trace analysis.
- 62Solvent extraction, activated charcoal, and electrolytic decomposition are methods used for analyte separation and_____.
Hint
These procedures can increase the amount of analyte available for measurement.
Answer
enrichment
Explanation
The listed techniques are used to separate and enrich analytes.
- 63Chelating agents such as EDTA and NTA bind many metals to form_____.
Hint
The metal becomes associated with a ligand before extraction.
Answer
complexes
Explanation
Chelating agents complex metals over a wide pH range before solvent extraction.
- 64After chelation, metals may be extracted into methyl ethyl ketone, abbreviated_____.
Hint
The source gives this abbreviation for the named solvent.
Answer
MBK
Explanation
The source describes extraction into MBK, or methyl ethyl ketone, or another suitable solvent.
- 65Preconcentration with_____is a well-known example of analyte enrichment.
Hint
This porous material can collect analyte from a sample.
Answer
activated charcoal
Explanation
Activated charcoal preconcentration is given as a familiar method for enriching an analyte.
- 66Some samples must be dissolved in acids under a_____atmosphere.
Hint
The digestion takes place under pressure rather than at ordinary atmospheric pressure.
Answer
pressurized
Explanation
The source notes that samples sometimes need acid dissolution under a pressurized atmosphere.
- 67For trace-level work, the analysis should use ultra-pure water and_____.
Hint
These reagents are used in sample preparation and must not add contaminants.
Answer
acids
Explanation
Ultra-pure water and acids are needed to reduce contamination in analyses at very low concentration levels.
- 68Biological analysis may measure calcium, magnesium, sodium, potassium, and_____in body fluids and tissues.
Hint
This nutrient metal is included in the listed elements of interest.
Answer
zinc
Explanation
The source lists calcium, magnesium, sodium, potassium, and zinc for analysis in body fluids and tissues.
- 69Before analysis, tissue samples may require wet or dry_____.
Hint
This sample-preparation process removes organic material by oxidation.
Answer
ashing
Explanation
The source states that tissue samples must be wet or dry ashed before analysis.
- 70Aluminum in brain tissue has received attention because of its possible connection with_____disease.
Hint
This is a neurodegenerative disease mentioned in the biological applications.
Answer
Alzheimer's
Explanation
The source notes that aluminum in brain tissue is a focus of study owing to Alzheimer's disease.
- 71Arsenic, mercury, cadmium, selenium, and thallium are important elements in_____applications.
Hint
These applications include analysis of body fluids and tissues.
Answer
biological
Explanation
The source identifies these potentially important elements in biological samples.
- 72Many dairy products, including ghee, butter, and cheese, must be wet_____before analysis.
Hint
This is the digestion treatment used for many organic food samples.
Answer
ashed
Explanation
The source states that almost all dairy products must be wet ashed before analysis.
- 73A common contaminant in foodstuffs that receives particular analytical attention is_____.
Hint
This toxic heavy metal is emphasized in food testing.
Answer
lead
Explanation
The source describes lead as a common food contaminant and highlights its analysis.
- 74Tin may be added to_____foils.
Hint
These packaging materials are made from a synthetic polymer.
Answer
plastic
Explanation
The source notes that tin is added to plastic foils.
- 75Soil analysis by ICP-AES can include sodium, potassium, calcium, magnesium, copper, manganese, zinc, boron, arsenic, cadmium, and_____.
Hint
This toxic element is also monitored as a soil contaminant.
Answer
lead
Explanation
Lead is included in the source's list of elements important in soil, fertilizer, and plant analysis.
- 76In water analysis, APDC and MIBK may be used to concentrate elements of_____.
Hint
These are the target elements selected for measurement.
Answer
interest
Explanation
The source describes using APDC and MIBK to concentrate elements of interest in water.
- 77Sediment digestion may require HF and_____because the samples can be difficult to dissolve.
Hint
This is a strong acid mixture used to dissolve resistant materials.
Answer
aqua regia
Explanation
The source identifies HF and aqua regia as reagents used for difficult sediment digestion.
- 78The source describes the environment as a system in which air, water, soil, organisms, and human activities_____one another.
Hint
The components do not act independently.
Answer
influence
Explanation
Environmental factors interact mutually, so it is difficult to isolate one factor completely.
- 79A large share of particulate matter in air originates from natural sources such as_____and aerosols.
Hint
This naturally occurring airborne material can enter the respiratory system.
Answer
dust
Explanation
The source identifies dust and aerosols as natural sources of airborne particulate matter.
- 80Airborne particles around 2.5 microns in size can enter and deposit in the_____.
Hint
These are the organs affected when fine particles are inhaled.
Answer
lungs
Explanation
The source states that aerosols of about 2.5 microns can enter and deposit in the lungs.
- 81The source notes that stack gases can carry particulate matter as far as_____kilometers.
Hint
This is the approximate maximum distance stated for transported particles.
Answer
400
Explanation
Particulate matter from stack gases can be carried up to about 400 kilometers.
- 82Cadmium and lead in flue dust can be collected using_____techniques.
Hint
The particles are physically captured from the treated gas stream.
Answer
filtration
Explanation
The source describes filtration as a method for collecting cadmium and lead from flue dust.
- 83Filtration sampling of flue dust may last four, six, or_____hours depending on requirements.
Hint
This is the longest sampling duration listed.
Answer
24
Explanation
The source gives four, six, and twenty-four hours as possible collection periods.
- 84Toxic heavy metals, including mercury, can be transferred to soils and_____water.
Hint
This is water found above ground, as opposed to subterranean water.
Answer
surface
Explanation
The source notes that toxic heavy metals can pass into soils and surface water.
- 85Sewage sludge and electronic commodities require_____before elemental analysis.
Hint
The samples must be broken down or brought into a suitable solution.
Answer
digestion
Explanation
The source states that sewage sludge and electronic commodities require digestion.
- 86ICP was developed mainly for analyzing rocks, minerals, and_____.
Hint
These are naturally occurring materials mined for their useful constituents.
Answer
ores
Explanation
The source describes the original main use of ICP as analysis of rocks, minerals, and ores.
- 87For mineral digestion, the source identifies HF and_____acid as a preferred combination.
Hint
This strong mineral acid is paired with HF in the described method.
Answer
sulfuric
Explanation
Digestion with HF and sulfuric acid is described as a preferred technique for minerals.
- 88Volatile elements such as mercury and cadmium may require pressurization using a_____bomb.
Hint
This is the name given to the pressure vessel in the source.
Answer
power
Explanation
The source recommends pressurization using a power bomb for volatile elements such as mercury and cadmium.
- 89A common fusion procedure uses sodium carbonate and sodium bromide, followed by dissolving the melt in_____.
Hint
The fused material is brought into solution using an acidic medium.
Answer
acid
Explanation
The source describes fusion with sodium carbonate and sodium bromide, then dissolution of the melt in acid.
- 90Gold is usually extracted and enriched before measurement because its concentration is too low for_____analysis.
Hint
The alternative would measure the untreated sample without enrichment.
Answer
direct
Explanation
The source explains that gold is commonly enriched before determination because direct analysis is not sensitive enough for its low concentration.
- 91In metallurgy and plating, ICP-AES is used to determine metals such as aluminum, vanadium, titanium, chromium, and_____.
Hint
This transition metal completes the list of examples.
Answer
niobium
Explanation
The source lists niobium along with aluminum, vanadium, titanium, and chromium for metallurgical analysis.
- 92For plating solutions, treatment with hydrochloric acid and_____is described as suitable.
Hint
The paired acid is represented in the source by the abbreviation HNO3.
Answer
nitric acid
Explanation
The source identifies hydrochloric and nitric acid treatment as suitable for plating solutions.
- 93ICP-AES can determine trace bismuth, lead, selenium, tellurium, and thallium in complex_____alloys.
Hint
The alloy is based on a metal often used for corrosion-resistant materials.
Answer
nickel
Explanation
The source identifies complex nickel alloys as matrices in which these trace elements can be measured.
- 94An aluminum alloy can be dissolved for analysis in one-to-one hydrochloric acid with a few drops of_____.
Hint
This oxidizing reagent is added to the acid treatment.
Answer
hydrogen peroxide
Explanation
The described aluminum-alloy preparation uses a one-to-one hydrochloric acid solution with a few drops of hydrogen peroxide.
- 95Iron and copper are monitored in lubricating oil because they are used as_____.
Hint
They are intentionally included components rather than necessarily contaminants.
Answer
additives
Explanation
The source notes that iron and copper additives in lubricating oil are regularly monitored at ppm quantities.
- 96Lead in petrol is a frequently studied topic for_____analysts.
Hint
These analysts focus on pollutants and their effects outside the laboratory.
Answer
environmental
Explanation
The source describes lead in petrol as a favorite topic of environmental analysts.
- 97Examples of materials analyzed for metals in the plastics, paper, and textiles sector include polypropylene, nylon 66, polystyrene, and_____.
Hint
This common polymer abbreviation is listed with the other materials.
Answer
PVC
Explanation
The source lists PVC among polymers considered in this area of analysis.
- 98The listed polymers can be dissolved in MIBK and_____directly when suitable.
Hint
The prepared solution is introduced into the instrument through the sample inlet.
Answer
aspirated
Explanation
The source describes dissolving certain polymers in MIBK and aspirating them directly into ICP-AES.
- 99Insoluble polymers must be digested and brought into_____before analysis.
Hint
The sample must be made into a liquid form suitable for introduction.
Answer
solution
Explanation
The source states that insoluble polymers need digestion to bring them into solution.
- 100Organotin compounds may be present in packing materials as_____.
Hint
They are added to help maintain the material's properties.
Answer
stabilizers
Explanation
The source notes that packing materials may contain organotin compounds used as stabilizers.
- 101Metal-plastic samples may be extracted with three percent_____acid.
Hint
This weak organic acid is used for the stated extraction.
Answer
acetic
Explanation
The source gives three percent acetic acid as an extraction medium for metal-plastic samples.
- 102Matrix matching is important in ICP-AES because the sample matrix can affect measurements at_____concentrations.
Hint
The concern is especially serious when analyte amounts are very small.
Answer
low
Explanation
The source stresses matrix matching because otherwise elements at very low concentration may not be determined reliably.
- 103In paper, pulp, and cellulose, direct determinations have included cadmium, copper, lead, and_____.
Hint
This metal is also listed in the examples of polymer and material analysis.
Answer
manganese
Explanation
The source lists manganese with cadmium, copper, and lead for direct determination in paper-related matrices.
- 104Pharmaceutical and miscellaneous product analysis may include sodium, potassium, calcium, arsenic, cobalt, manganese, and_____.
Hint
This precious metal is the final example in the source's list.
Answer
silver
Explanation
The source includes silver among elements of importance in pharmaceuticals and other products.
- 105Cobalt in vitamin B12 and proteins can be measured using flame AAS or_____.
Hint
This is the atomic emission technique discussed throughout the content.
Answer
ICP-AES
Explanation
The source states that cobalt in vitamin B12 and proteins can be determined by flame AAS as well as ICP-AES.
- 106Dietary supplements are routinely checked for arsenic and_____.
Hint
This element is also important in biological and environmental analysis.
Answer
selenium
Explanation
The source identifies arsenic and selenium as elements routinely checked in dietary supplements.
- 107Uranium in yellow cake is usually determined by_____.
Hint
This is the emission spectrometry technique used for the measurement.
Answer
ICP-AES
Explanation
The source states that uranium in yellow cake is usually determined using ICP-AES.
- 108A typical ICP-AES workflow begins by preparing the samples and_____.
Hint
Known solutions are needed alongside samples for calibration.
Answer
standards
Explanation
The first step in the described workflow is preparing samples and standards.
- 109If a ready-made method is available, the operator recalls it in the_____.
Hint
The instrument's control system stores and loads methods.
Answer
computer
Explanation
The described workflow has the operator tell the computer to recall an existing method.
- 110Before running samples, calibrated samples are aspirated using_____solutions.
Hint
These known solutions establish or check the instrument response.
Answer
standard
Explanation
The normal workflow aspirates calibrated samples using standard solutions before running the samples.
- 111When no ready-made method exists, the operator selects elements, collects representative spectra, and chooses an_____time.
Hint
This parameter controls the period over which signal is collected.
Answer
integration
Explanation
Method development includes selecting integration time, data acquisition mode, and standard concentrations.
- 112When developing a new method, the operator selects the background and then_____the method.
Hint
The selected settings need to be saved for later use.
Answer
stores
Explanation
The source describes selecting the background and storing the method during method development.
- 113After samples are run in the normal ICP-AES scheme, the instrument is used to generate the_____.
Hint
This is the document presenting the analysis results.
Answer
report
Explanation
The described sequence ends by generating a report after running the samples.