Atomic Emission Spectrometry Sources

  • 113Unfamiliar
  • 0Learning
  • 0Familiar
  • 0Mastered
Topic: Source Requirements
For an ideal atomic emission source, the plasma must be hot enough to excite every analyte element to the
.

113 questions

  1. 1
    For an ideal atomic emission source, the plasma must be hot enough to excite every analyte element to the
    _____
    .
    Hint

    Think about the energy transition required for emission.

    Answer

    next higher energy level

    Explanation

    A sufficiently high temperature is needed to excite all elements being analyzed to the next higher energy level.

  2. 2
    A stable plasma is important because the source temperature should remain
    _____
    .
    Hint

    Consider whether temperature fluctuations would support reliable excitation.

    Answer

    stable

    Explanation

    The ideal source should have a very stable plasma so its temperature and excitation conditions remain consistent.

  3. 3
    To improve detectability, the ideal source should have background intensity that is as
    _____
    as possible.
    Hint

    A weaker background makes analyte signals easier to distinguish.

    Answer

    low

    Explanation

    Low background intensity improves the detectability of analyte emission.

  4. 4
    A uniform temperature and low optical density help produce
    _____
    calibration curves.
    Hint

    These source properties support a proportional signal response.

    Answer

    linear

    Explanation

    The source specifications associate uniform temperature and low optical density with linear calibration curves.

  5. 5
    The ideal source should accept representative subsamples at a
    _____
    rate.
    Hint

    The sample throughput should be practical rather than excessively slow.

    Answer

    reasonable

    Explanation

    The source should allow a representative subsample to be taken and introduced in an acceptable amount of time.

  6. 6
    An ideal source should be able to accept samples in solid, liquid, or
    _____
    form.
    Hint

    This is the third common state of matter listed with solids and liquids.

    Answer

    gaseous

    Explanation

    The ideal source should accommodate samples in solid, liquid, or gaseous forms.

  7. 7
    Introducing a sample should not drastically alter the source's
    _____
    .
    Hint

    The relevant property describes the energy state within the source.

    Answer

    internal energy

    Explanation

    An ideal source remains stable when a sample is introduced, without a drastic change to its internal energy.

  8. 8
    Ideally, all sample material entering the source should be converted into
    _____
    .
    Hint

    The desired products are uncombined atoms that can produce atomic emission.

    Answer

    free atoms

    Explanation

    The ideal source converts all of the sample material into free atoms.

  9. 9
    The ideal source should use commonly available, reasonably priced
    _____
    .
    Hint

    These are used to operate the plasma source.

    Answer

    gases

    Explanation

    Commonly available and reasonably priced operating gases are a desired source property.

  10. 10
    In the described ICP, the commonly used plasma gas is
    _____
    .
    Hint

    It is a noble gas used to sustain the plasma.

    Answer

    argon

    Explanation

    The source describes the ICP as operating on argon, which is reasonably priced and readily available.

  11. 11
    The ICP comes close to fulfilling the specifications for an
    _____
    atomic emission source.
    Hint

    This is the standard against which the source properties are compared.

    Answer

    ideal

    Explanation

    The ICP is presented as coming close to the desired properties of an ideal source.

  12. 12
    The ICP is optically almost transparent, which helps keep its
    _____
    low.
    Hint

    A low value of this signal makes analyte emission easier to detect.

    Answer

    background

    Explanation

    The ICP's near transparency is given as a reason for its very low background.

  13. 13
    The ICP tail flame is described as optically thin, so it produces little
    _____
    .
    Hint

    The low signal from the tail flame reduces the unwanted baseline.

    Answer

    emission

    Explanation

    The optically thin tail flame contributes little emission, keeping background low.

  14. 14
    The sample is introduced into the ICP at a reasonable rate using a
    _____
    pump.
    Hint

    This pump type moves liquid by compressing and releasing flexible tubing.

    Answer

    peristaltic

    Explanation

    The described ICP sample input uses a peristaltic pump to deliver the sample at a reasonable rate.

  15. 15
    ICP-AES allows rapid and
    _____
    analysis of multiple elements.
    Hint

    Several elements can be measured during the same analysis.

    Answer

    simultaneous

    Explanation

    Rapid simultaneous multi-element analysis is listed as an advantage of ICP-AES.

  16. 16
    The ICP is an excellent atomic emission source in part because it is optically
    _____
    .
    Hint

    This property limits self-absorption and background effects.

    Answer

    thin

    Explanation

    The source identifies the ICP as an excellent emission source because it is optically thin.

  17. 17
    The stated ICP-AES detection limits are about one to one hundred nanograms per gram or micrograms per liter, corresponding to
    _____
    .
    Hint

    This concentration scale is abbreviated ppb.

    Answer

    parts per billion

    Explanation

    The described detection limits are in the parts-per-billion range.

  18. 18
    ICP-AES is suitable for measuring concentrations from trace levels through
    _____
    levels.
    Hint

    This is the high-concentration end of the stated range.

    Answer

    major

    Explanation

    The technique is described as suitable across concentrations from trace to major levels.

  19. 19
    The performance of ICP-AES is described as having very good accuracy and
    _____
    .
    Hint

    This property concerns the agreement among repeated measurements.

    Answer

    precision

    Explanation

    Very good accuracy and precision are both listed among the advantages of ICP-AES.

  20. 20
    In principle, ICP-AES can determine all elements except
    _____
    .
    Hint

    This is also the gas used to operate the plasma.

    Answer

    argon

    Explanation

    The source states that ICP-AES can in principle be used for all elements except the plasma gas argon.

  21. 21
    In practice, about
    _____
    elements can be determined using a single ICP-AES plasma source.
    Hint

    The source gives an approximate count rather than a complete list.

    Answer

    70

    Explanation

    The described practical capability is approximately seventy elements using one plasma source.

  22. 22
    Elements that can be difficult to determine by AAS but are readily determined by ICP-AES include boron, carbon, and
    _____
    .
    Hint

    Cerium and lanthanum are examples of this group.

    Answer

    rare earth elements

    Explanation

    The lecture gives boron and carbon as examples and also mentions rare earth elements such as cerium and lanthanum.

  23. 23
    Titanium, tantalum, vanadium, and zirconium are examples of
    _____
    elements that ICP-AES can determine.
    Hint

    These elements are characterized by resistance to heat and difficult processing.

    Answer

    refractory

    Explanation

    The source lists titanium, tantalum, vanadium, and zirconium as refractory elements that can be determined by ICP-AES.

  24. 24
    ICP calibration graphs are essentially linear below 10 ppm and can extend up to about
    _____
    ppm.
    Hint

    This upper value is stated as one thousand parts per million.

    Answer

    1000

    Explanation

    The described wide dynamic range includes essentially linear calibration below 10 ppm and up to about 1000 ppm.

  25. 25
    When sensitivity is low, even the
    _____
    solution can sometimes be measured directly.
    Hint

    It is the concentrated solution from which dilutions are prepared.

    Answer

    stock

    Explanation

    The wide dynamic range may allow the stock solution itself to be measured if the sensitivity is low.

  26. 26
    A solution prepared suitably from the test portion submitted for analysis is called the
    _____
    solution.
    Hint

    It is the prepared solution that is aspirated into the ICP.

    Answer

    sample

    Explanation

    The sample solution is made from the test portion and is the solution presented for analysis.

  27. 27
    A solution of known concentration used to prepare dilutions for calibration standards is a
    _____
    solution.
    Hint

    It serves as the starting solution for making standards.

    Answer

    stock

    Explanation

    A stock solution has a known concentration and is used to prepare diluted calibration standards.

  28. 28
    A series of solutions with known analyte concentrations in a solvent is called
    _____
    solutions.
    Hint

    These known solutions provide comparison values for analysis.

    Answer

    reference

    Explanation

    Reference solutions are a series of known analyte concentrations in a solvent.

  29. 29
    Reference solutions may contain the analyte and also the sample's
    _____
    elements.
    Hint

    These are the other components surrounding the analyte in the sample.

    Answer

    matrix

    Explanation

    Reference solutions can include matrix elements in addition to the analyte.

  30. 30
    A blank solution contains the matrix components but has zero concentration of the
    _____
    .
    Hint

    This is the element whose concentration is being measured.

    Answer

    analyte

    Explanation

    The blank is a matrix-matched solution that does not contain the analyte.

  31. 31
    A graphical plot of measured absorbance or emission against analyte concentration or mass is the
    _____
    .
    Hint

    It is also called the calibration curve.

    Answer

    analytical curve

    Explanation

    The analytical curve relates measured absorbance or emission to the concentration or mass of the analyte.

  32. 32
    The concentration or mass corresponding to a net signal of one percent, or absorbance of 0.044, is called the
    _____
    .
    Hint

    This term identifies a signal response at a specified small level.

    Answer

    characteristic concentration or characteristic mass

    Explanation

    Characteristic concentration or characteristic mass refers to the analyte level corresponding to the stated net signal or absorbance.

  33. 33
    When integrated absorbance or peak area is used, the absorbance specified for characteristic concentration is
    _____
    .
    Hint

    It is the numerical absorbance value stated alongside a net one-percent signal.

    Answer

    0.044

    Explanation

    The source gives 0.044 absorbance as the characteristic value when integrated absorbance or peak area is used.

  34. 34
    The slope of the analytical curve is called its
    _____
    .
    Hint

    It describes how much the signal changes as analyte amount changes.

    Answer

    sensitivity

    Explanation

    Sensitivity is defined as the slope of the analytical curve.

  35. 35
    Good sensitivity means a large change in signal for a
    _____
    change in concentration.
    Hint

    A method is more responsive when little analyte change produces a strong signal change.

    Answer

    small

    Explanation

    Good sensitivity requires the signal to change substantially when concentration changes only slightly.

  36. 36
    Accuracy describes how closely a measured mean agrees with the
    _____
    value.
    Hint

    This is the value the analytical result aims to estimate.

    Answer

    true

    Explanation

    Accuracy concerns the closeness of agreement between the true analyte value and the mean measured value.

  37. 37
    Precision describes the agreement among results obtained by repeating the analytical procedure under the same
    _____
    .
    Hint

    Keep the procedure and measurement setup consistent in the repeated analyses.

    Answer

    conditions

    Explanation

    Precision is the closeness of agreement among repeated results obtained using the same analytical procedure.

  38. 38
    Repeated results can be precise without being accurate if they all contain the same
    _____
    .
    Hint

    A consistent systematic problem can make results agree with each other but miss the true value.

    Answer

    error

    Explanation

    Precision does not guarantee accuracy because the same error may be present in every repeated measurement.

  39. 39
    The standard deviation describes the spread of measurements around their
    _____
    .
    Hint

    It is the central average of the measurement set.

    Answer

    mean value

    Explanation

    Standard deviation characterizes how measurements are distributed around their mean.

  40. 40
    Relative standard deviation is calculated by dividing standard deviation by the
    _____
    .
    Hint

    It scales the spread relative to the central measurement.

    Answer

    mean value

    Explanation

    Relative standard deviation is the standard deviation divided by the mean value.

  41. 41
    The lowest concentration measurable with the prescribed precision in a practical procedure is the
    _____
    .
    Hint

    At this level, the analyte is quantitatively determined rather than merely noticed.

    Answer

    determination limit

    Explanation

    The determination limit is the lowest concentration that can be measured with the prescribed precision.

  42. 42
    The detection limit is calculated from signal sensitivity and the absolute standard deviation, multiplied by a factor
    _____
    .
    Hint

    The source assigns this factor a value of two or three.

    Answer

    k

    Explanation

    The detection-limit expressions use a factor called k together with absolute standard deviation and signal sensitivity.

  43. 43
    In the detection-limit calculation, the factor k is usually
    _____
    or three.
    Hint

    The two possible values are small whole numbers.

    Answer

    2

    Explanation

    The stated value of k is two or three.

  44. 44
    The sigma used in the detection-limit calculation represents the
    _____
    standard deviation.
    Hint

    It is not the relative standard deviation.

    Answer

    absolute

    Explanation

    Sigma in the detection-limit expression is the absolute standard deviation.

  45. 45
    A detection limit may be measured using either peak area or peak
    _____
    .
    Hint

    This is the other common measure of an emission-line signal.

    Answer

    height

    Explanation

    Detection limits can be determined from either peak area or peak height.

  46. 46
    The detection limit depends on electrical noise, optical noise, sensitivity, analyte metastable energy status, and the chemical
    _____
    .
    Hint

    This includes the matrix surrounding the analyte.

    Answer

    environment

    Explanation

    These factors, including the analyte's chemical environment and matrix matching, affect the detection limit.

  47. 47
    Because detection limits vary among systems, measurement reliability depends on factors such as standard deviation, relative standard deviation, and
    _____
    .
    Hint

    This is the response of signal to a change in analyte concentration or mass.

    Answer

    sensitivity

    Explanation

    Reliability depends on several measurable characteristics, including detection limit, standard deviations, slope, and sensitivity.

  48. 48
    A signal around three times the mean background signal is used as an estimate of the limit of
    _____
    .
    Hint

    At this threshold, the element may be present but is not necessarily confirmed quantitatively.

    Answer

    detection

    Explanation

    The illustrated limit of detection is approximately three times the mean background signal.

  49. 49
    A signal higher than the detection threshold that confirms the element's presence is called the limit of
    _____
    .
    Hint

    This threshold provides stronger evidence that the analyte is present.

    Answer

    identification

    Explanation

    The limit of identification is a higher signal level used to make sure the element is present.

  50. 50
    The threshold at which the amount of an analyte can be measured quantitatively is the limit of
    _____
    .
    Hint

    This is the level at which determining how much analyte is present becomes possible.

    Answer

    determination

    Explanation

    Above the limit of determination, quantitative measurement is possible with the prescribed precision.

  51. 51
    For thirty or more measurements, the described precision estimate uses standard deviation multiplied by
    _____
    .
    Hint

    This is the numerical multiplier specified for large measurement sets.

    Answer

    2.83

    Explanation

    The source gives 2.83 times the standard deviation for thirty or more measurements.

  52. 52
    For other measurement counts, the described precision estimate can use t multiplied by the standard deviation, where t is the
    _____
    factor.
    Hint

    This factor is named after a scientist and depends on the statistical context.

    Answer

    Student's

    Explanation

    The source refers to t as the Student's factor.

  53. 53
    In a standard deviation calculation, the number of measurements is represented by
    _____
    .
    Hint

    It is the count of observations in the measurement set.

    Answer

    n

    Explanation

    The source identifies n as the number of measurements used to calculate standard deviation.

  54. 54
    The average of all measurements in a standard deviation calculation is the
    _____
    .
    Hint

    It is the central value around which individual results vary.

    Answer

    mean value

    Explanation

    The mean value is the average of all measurements and is used to describe their spread.

  55. 55
    AAS is described as a relative technique rather than an
    _____
    technique.
    Hint

    Its quantitative results are obtained by comparison rather than independently.

    Answer

    absolute

    Explanation

    AAS is a relative technique: quantitative results come from comparison with reference solutions or materials.

  56. 56
    Because AAS is a relative technique, quantitative analysis requires comparison with reference solutions or reference
    _____
    .
    Hint

    These provide known values for comparison with the sample.

    Answer

    materials

    Explanation

    Reference measurements must be made using reference solutions or reference materials.

  57. 57
    The analytical curve, bracketing, and analyte addition are all calibration
    _____
    .
    Hint

    They are different ways to establish or apply a concentration response.

    Answer

    techniques

    Explanation

    The source lists the analytical curve, bracketing, and analyte addition as techniques for determining sample concentrations.

  58. 58
    In AES, the use of
    _____
    standards is identified as a calibration need.
    Hint

    These standards are introduced to help account for variation during measurement.

    Answer

    internal

    Explanation

    The source states that internal standards need to be employed in AES.

  59. 59
    When ICP sensitivity is insufficient for trace analysis, a more
    _____
    method may be needed.
    Hint

    The alternative must respond more strongly to small analyte quantities.

    Answer

    sensitive

    Explanation

    Insufficient sensitivity can require a more sensitive analytical method or a sample enrichment step.

  60. 60
    In trace and ultra-trace analysis, every sample pretreatment is a potential source of
    _____
    .
    Hint

    Even small amounts introduced during handling can affect results.

    Answer

    contamination

    Explanation

    Sample pretreatment can introduce contamination, so it should be minimized and carefully controlled.

  61. 61
    One way to reduce contamination risk is to introduce the sample as it is and avoid unnecessary
    _____
    .
    Hint

    This refers to processing the sample before measurement.

    Answer

    pretreatment

    Explanation

    The source emphasizes avoiding sample pretreatment whenever possible in trace and ultra-trace analysis.

  62. 62
    Solvent extraction, activated charcoal, and electrolytic decomposition are methods used for analyte separation and
    _____
    .
    Hint

    These procedures can increase the amount of analyte available for measurement.

    Answer

    enrichment

    Explanation

    The listed techniques are used to separate and enrich analytes.

  63. 63
    Chelating agents such as EDTA and NTA bind many metals to form
    _____
    .
    Hint

    The metal becomes associated with a ligand before extraction.

    Answer

    complexes

    Explanation

    Chelating agents complex metals over a wide pH range before solvent extraction.

  64. 64
    After chelation, metals may be extracted into methyl ethyl ketone, abbreviated
    _____
    .
    Hint

    The source gives this abbreviation for the named solvent.

    Answer

    MBK

    Explanation

    The source describes extraction into MBK, or methyl ethyl ketone, or another suitable solvent.

  65. 65
    Preconcentration with
    _____
    is a well-known example of analyte enrichment.
    Hint

    This porous material can collect analyte from a sample.

    Answer

    activated charcoal

    Explanation

    Activated charcoal preconcentration is given as a familiar method for enriching an analyte.

  66. 66
    Some samples must be dissolved in acids under a
    _____
    atmosphere.
    Hint

    The digestion takes place under pressure rather than at ordinary atmospheric pressure.

    Answer

    pressurized

    Explanation

    The source notes that samples sometimes need acid dissolution under a pressurized atmosphere.

  67. 67
    For trace-level work, the analysis should use ultra-pure water and
    _____
    .
    Hint

    These reagents are used in sample preparation and must not add contaminants.

    Answer

    acids

    Explanation

    Ultra-pure water and acids are needed to reduce contamination in analyses at very low concentration levels.

  68. 68
    Biological analysis may measure calcium, magnesium, sodium, potassium, and
    _____
    in body fluids and tissues.
    Hint

    This nutrient metal is included in the listed elements of interest.

    Answer

    zinc

    Explanation

    The source lists calcium, magnesium, sodium, potassium, and zinc for analysis in body fluids and tissues.

  69. 69
    Before analysis, tissue samples may require wet or dry
    _____
    .
    Hint

    This sample-preparation process removes organic material by oxidation.

    Answer

    ashing

    Explanation

    The source states that tissue samples must be wet or dry ashed before analysis.

  70. 70
    Aluminum in brain tissue has received attention because of its possible connection with
    _____
    disease.
    Hint

    This is a neurodegenerative disease mentioned in the biological applications.

    Answer

    Alzheimer's

    Explanation

    The source notes that aluminum in brain tissue is a focus of study owing to Alzheimer's disease.

  71. 71
    Arsenic, mercury, cadmium, selenium, and thallium are important elements in
    _____
    applications.
    Hint

    These applications include analysis of body fluids and tissues.

    Answer

    biological

    Explanation

    The source identifies these potentially important elements in biological samples.

  72. 72
    Many dairy products, including ghee, butter, and cheese, must be wet
    _____
    before analysis.
    Hint

    This is the digestion treatment used for many organic food samples.

    Answer

    ashed

    Explanation

    The source states that almost all dairy products must be wet ashed before analysis.

  73. 73
    A common contaminant in foodstuffs that receives particular analytical attention is
    _____
    .
    Hint

    This toxic heavy metal is emphasized in food testing.

    Answer

    lead

    Explanation

    The source describes lead as a common food contaminant and highlights its analysis.

  74. 74
    Tin may be added to
    _____
    foils.
    Hint

    These packaging materials are made from a synthetic polymer.

    Answer

    plastic

    Explanation

    The source notes that tin is added to plastic foils.

  75. 75
    Soil analysis by ICP-AES can include sodium, potassium, calcium, magnesium, copper, manganese, zinc, boron, arsenic, cadmium, and
    _____
    .
    Hint

    This toxic element is also monitored as a soil contaminant.

    Answer

    lead

    Explanation

    Lead is included in the source's list of elements important in soil, fertilizer, and plant analysis.

  76. 76
    In water analysis, APDC and MIBK may be used to concentrate elements of
    _____
    .
    Hint

    These are the target elements selected for measurement.

    Answer

    interest

    Explanation

    The source describes using APDC and MIBK to concentrate elements of interest in water.

  77. 77
    Sediment digestion may require HF and
    _____
    because the samples can be difficult to dissolve.
    Hint

    This is a strong acid mixture used to dissolve resistant materials.

    Answer

    aqua regia

    Explanation

    The source identifies HF and aqua regia as reagents used for difficult sediment digestion.

  78. 78
    The source describes the environment as a system in which air, water, soil, organisms, and human activities
    _____
    one another.
    Hint

    The components do not act independently.

    Answer

    influence

    Explanation

    Environmental factors interact mutually, so it is difficult to isolate one factor completely.

  79. 79
    A large share of particulate matter in air originates from natural sources such as
    _____
    and aerosols.
    Hint

    This naturally occurring airborne material can enter the respiratory system.

    Answer

    dust

    Explanation

    The source identifies dust and aerosols as natural sources of airborne particulate matter.

  80. 80
    Airborne particles around 2.5 microns in size can enter and deposit in the
    _____
    .
    Hint

    These are the organs affected when fine particles are inhaled.

    Answer

    lungs

    Explanation

    The source states that aerosols of about 2.5 microns can enter and deposit in the lungs.

  81. 81
    The source notes that stack gases can carry particulate matter as far as
    _____
    kilometers.
    Hint

    This is the approximate maximum distance stated for transported particles.

    Answer

    400

    Explanation

    Particulate matter from stack gases can be carried up to about 400 kilometers.

  82. 82
    Cadmium and lead in flue dust can be collected using
    _____
    techniques.
    Hint

    The particles are physically captured from the treated gas stream.

    Answer

    filtration

    Explanation

    The source describes filtration as a method for collecting cadmium and lead from flue dust.

  83. 83
    Filtration sampling of flue dust may last four, six, or
    _____
    hours depending on requirements.
    Hint

    This is the longest sampling duration listed.

    Answer

    24

    Explanation

    The source gives four, six, and twenty-four hours as possible collection periods.

  84. 84
    Toxic heavy metals, including mercury, can be transferred to soils and
    _____
    water.
    Hint

    This is water found above ground, as opposed to subterranean water.

    Answer

    surface

    Explanation

    The source notes that toxic heavy metals can pass into soils and surface water.

  85. 85
    Sewage sludge and electronic commodities require
    _____
    before elemental analysis.
    Hint

    The samples must be broken down or brought into a suitable solution.

    Answer

    digestion

    Explanation

    The source states that sewage sludge and electronic commodities require digestion.

  86. 86
    ICP was developed mainly for analyzing rocks, minerals, and
    _____
    .
    Hint

    These are naturally occurring materials mined for their useful constituents.

    Answer

    ores

    Explanation

    The source describes the original main use of ICP as analysis of rocks, minerals, and ores.

  87. 87
    For mineral digestion, the source identifies HF and
    _____
    acid as a preferred combination.
    Hint

    This strong mineral acid is paired with HF in the described method.

    Answer

    sulfuric

    Explanation

    Digestion with HF and sulfuric acid is described as a preferred technique for minerals.

  88. 88
    Volatile elements such as mercury and cadmium may require pressurization using a
    _____
    bomb.
    Hint

    This is the name given to the pressure vessel in the source.

    Answer

    power

    Explanation

    The source recommends pressurization using a power bomb for volatile elements such as mercury and cadmium.

  89. 89
    A common fusion procedure uses sodium carbonate and sodium bromide, followed by dissolving the melt in
    _____
    .
    Hint

    The fused material is brought into solution using an acidic medium.

    Answer

    acid

    Explanation

    The source describes fusion with sodium carbonate and sodium bromide, then dissolution of the melt in acid.

  90. 90
    Gold is usually extracted and enriched before measurement because its concentration is too low for
    _____
    analysis.
    Hint

    The alternative would measure the untreated sample without enrichment.

    Answer

    direct

    Explanation

    The source explains that gold is commonly enriched before determination because direct analysis is not sensitive enough for its low concentration.

  91. 91
    In metallurgy and plating, ICP-AES is used to determine metals such as aluminum, vanadium, titanium, chromium, and
    _____
    .
    Hint

    This transition metal completes the list of examples.

    Answer

    niobium

    Explanation

    The source lists niobium along with aluminum, vanadium, titanium, and chromium for metallurgical analysis.

  92. 92
    For plating solutions, treatment with hydrochloric acid and
    _____
    is described as suitable.
    Hint

    The paired acid is represented in the source by the abbreviation HNO3.

    Answer

    nitric acid

    Explanation

    The source identifies hydrochloric and nitric acid treatment as suitable for plating solutions.

  93. 93
    ICP-AES can determine trace bismuth, lead, selenium, tellurium, and thallium in complex
    _____
    alloys.
    Hint

    The alloy is based on a metal often used for corrosion-resistant materials.

    Answer

    nickel

    Explanation

    The source identifies complex nickel alloys as matrices in which these trace elements can be measured.

  94. 94
    An aluminum alloy can be dissolved for analysis in one-to-one hydrochloric acid with a few drops of
    _____
    .
    Hint

    This oxidizing reagent is added to the acid treatment.

    Answer

    hydrogen peroxide

    Explanation

    The described aluminum-alloy preparation uses a one-to-one hydrochloric acid solution with a few drops of hydrogen peroxide.

  95. 95
    Iron and copper are monitored in lubricating oil because they are used as
    _____
    .
    Hint

    They are intentionally included components rather than necessarily contaminants.

    Answer

    additives

    Explanation

    The source notes that iron and copper additives in lubricating oil are regularly monitored at ppm quantities.

  96. 96
    Lead in petrol is a frequently studied topic for
    _____
    analysts.
    Hint

    These analysts focus on pollutants and their effects outside the laboratory.

    Answer

    environmental

    Explanation

    The source describes lead in petrol as a favorite topic of environmental analysts.

  97. 97
    Examples of materials analyzed for metals in the plastics, paper, and textiles sector include polypropylene, nylon 66, polystyrene, and
    _____
    .
    Hint

    This common polymer abbreviation is listed with the other materials.

    Answer

    PVC

    Explanation

    The source lists PVC among polymers considered in this area of analysis.

  98. 98
    The listed polymers can be dissolved in MIBK and
    _____
    directly when suitable.
    Hint

    The prepared solution is introduced into the instrument through the sample inlet.

    Answer

    aspirated

    Explanation

    The source describes dissolving certain polymers in MIBK and aspirating them directly into ICP-AES.

  99. 99
    Insoluble polymers must be digested and brought into
    _____
    before analysis.
    Hint

    The sample must be made into a liquid form suitable for introduction.

    Answer

    solution

    Explanation

    The source states that insoluble polymers need digestion to bring them into solution.

  100. 100
    Organotin compounds may be present in packing materials as
    _____
    .
    Hint

    They are added to help maintain the material's properties.

    Answer

    stabilizers

    Explanation

    The source notes that packing materials may contain organotin compounds used as stabilizers.

  101. 101
    Metal-plastic samples may be extracted with three percent
    _____
    acid.
    Hint

    This weak organic acid is used for the stated extraction.

    Answer

    acetic

    Explanation

    The source gives three percent acetic acid as an extraction medium for metal-plastic samples.

  102. 102
    Matrix matching is important in ICP-AES because the sample matrix can affect measurements at
    _____
    concentrations.
    Hint

    The concern is especially serious when analyte amounts are very small.

    Answer

    low

    Explanation

    The source stresses matrix matching because otherwise elements at very low concentration may not be determined reliably.

  103. 103
    In paper, pulp, and cellulose, direct determinations have included cadmium, copper, lead, and
    _____
    .
    Hint

    This metal is also listed in the examples of polymer and material analysis.

    Answer

    manganese

    Explanation

    The source lists manganese with cadmium, copper, and lead for direct determination in paper-related matrices.

  104. 104
    Pharmaceutical and miscellaneous product analysis may include sodium, potassium, calcium, arsenic, cobalt, manganese, and
    _____
    .
    Hint

    This precious metal is the final example in the source's list.

    Answer

    silver

    Explanation

    The source includes silver among elements of importance in pharmaceuticals and other products.

  105. 105
    Cobalt in vitamin B12 and proteins can be measured using flame AAS or
    _____
    .
    Hint

    This is the atomic emission technique discussed throughout the content.

    Answer

    ICP-AES

    Explanation

    The source states that cobalt in vitamin B12 and proteins can be determined by flame AAS as well as ICP-AES.

  106. 106
    Dietary supplements are routinely checked for arsenic and
    _____
    .
    Hint

    This element is also important in biological and environmental analysis.

    Answer

    selenium

    Explanation

    The source identifies arsenic and selenium as elements routinely checked in dietary supplements.

  107. 107
    Uranium in yellow cake is usually determined by
    _____
    .
    Hint

    This is the emission spectrometry technique used for the measurement.

    Answer

    ICP-AES

    Explanation

    The source states that uranium in yellow cake is usually determined using ICP-AES.

  108. 108
    A typical ICP-AES workflow begins by preparing the samples and
    _____
    .
    Hint

    Known solutions are needed alongside samples for calibration.

    Answer

    standards

    Explanation

    The first step in the described workflow is preparing samples and standards.

  109. 109
    If a ready-made method is available, the operator recalls it in the
    _____
    .
    Hint

    The instrument's control system stores and loads methods.

    Answer

    computer

    Explanation

    The described workflow has the operator tell the computer to recall an existing method.

  110. 110
    Before running samples, calibrated samples are aspirated using
    _____
    solutions.
    Hint

    These known solutions establish or check the instrument response.

    Answer

    standard

    Explanation

    The normal workflow aspirates calibrated samples using standard solutions before running the samples.

  111. 111
    When no ready-made method exists, the operator selects elements, collects representative spectra, and chooses an
    _____
    time.
    Hint

    This parameter controls the period over which signal is collected.

    Answer

    integration

    Explanation

    Method development includes selecting integration time, data acquisition mode, and standard concentrations.

  112. 112
    When developing a new method, the operator selects the background and then
    _____
    the method.
    Hint

    The selected settings need to be saved for later use.

    Answer

    stores

    Explanation

    The source describes selecting the background and storing the method during method development.

  113. 113
    After samples are run in the normal ICP-AES scheme, the instrument is used to generate the
    _____
    .
    Hint

    This is the document presenting the analysis results.

    Answer

    report

    Explanation

    The described sequence ends by generating a report after running the samples.