Atomic Emission Spectrometry Sources
ICP-AES: Principles, Analytical Concepts, and Applications
1. Atomic emission spectrometry and the ideal source
Atomic emission spectrometry (AES) measures the light emitted by atoms or ions after they have been excited by an energy source. The emitted wavelengths identify the elements, while emission intensity is used to estimate their concentrations.
An ideal source for atomic emission analysis should have the following properties:
Sufficiently high temperature: The source must excite the elements being measured so that they emit detectable radiation. Its temperature should remain stable during analysis.
Low background emission: A low background makes weak analyte signals easier to distinguish and improves detectability.
Uniform temperature and low optical density: A uniform source promotes consistent excitation, while low optical density helps prevent emitted light from being absorbed or distorted within the source. These conditions support linear calibration curves.
Reasonable sample throughput: The source should accept representative sample portions at a practical rate.
Flexibility in sample form: Ideally, it should accommodate solid, liquid, and gaseous samples, either directly or after suitable preparation.
Minimal disturbance from sample introduction: Introducing a sample should not substantially change the source’s internal energy or stability.
Efficient atomization: Sample material should be converted into free atoms or ions that can be excited.
Practical operating gases: The source should use gases that are readily available at reasonable cost.
2. How ICP-AES approaches the ideal
Inductively coupled plasma atomic emission spectrometry (ICP-AES) uses a plasma, commonly sustained with argon, as its excitation source. It satisfies many of the requirements of an ideal source:
The plasma reaches temperatures high enough to excite most elements of analytical interest.
Its background is relatively low; the plasma is described as nearly optically transparent.
The temperature is fairly uniform, and the tail flame is optically thin, so it contributes relatively little unwanted emission.
Liquid samples can be introduced at a controlled, reasonable rate, commonly using a peristaltic pump.
The plasma is sufficiently stable that sample introduction does not usually cause a major change in its energy.
Liquid and gaseous samples can be converted into atoms or ions in the plasma. Solid samples generally require suitable sampling or conversion into a form the instrument can introduce.
Argon is widely available at reasonable cost.
The plasma is particularly useful because one source can excite many elements, allowing simultaneous multi-element analysis.
3. Advantages and capabilities of ICP-AES
ICP-AES has several important analytical advantages:
Rapid, simultaneous analysis: Many elements can be measured in a single sample run.
Optically thin source: The plasma’s optical properties help reduce self-absorption and support reliable emission measurements.
Broad linear range: Calibration curves may be linear across approximately four to six orders of magnitude, depending on the element, wavelength, and measurement conditions.
Low detection limits: Reported values can be in the range of approximately 1–100 ng/g or μg/L (parts per billion), although actual performance depends on the analytical system and sample matrix.
Wide concentration coverage: ICP-AES can be used from trace concentrations to major-element concentrations.
Good accuracy and precision: When calibration, sample preparation, and measurement conditions are appropriate, results can be both accurate and repeatable.
Broad elemental coverage: In principle, ICP-AES can determine nearly all elements other than argon, which is the plasma gas. In practice, roughly 70 elements are commonly measured with a single plasma source.
Ability to measure elements that can be difficult by atomic absorption spectrometry (AAS): Examples include boron, carbon, rare-earth elements such as cerium and lanthanum, actinide elements, and refractory elements such as titanium, tantalum, vanadium, and zirconium.
4. Sample solutions and calibration terminology
Many samples are analyzed as liquids, even when the original material is a solid. A solid may be digested, dissolved, or otherwise prepared before it is introduced into the plasma.
Sample solution
A sample solution is a solution prepared from a test portion of the submitted sample for analysis. It is the solution introduced into the ICP, commonly by aspiration.
Stock solution
A stock solution is a solution of known analyte concentration. It is used to prepare diluted solutions for calibration.
Reference solutions
Reference solutions are solutions containing known concentrations of the analyte. They are commonly prepared by diluting a stock solution. They may also contain selected matrix components when matrix matching is required.
Blank solution
A blank solution contains the solvent and relevant matrix components but no analyte. Its analyte concentration is zero. A matrix-matched blank helps distinguish the analyte signal from contributions caused by the solvent or other sample components.
Analytical curve
An analytical curve, also called a calibration curve, is a plot relating the measured response—such as emission intensity—to analyte concentration or mass. It is used to interpret the response from an unknown sample and estimate its analyte content.
5. Analytical performance terms
Characteristic concentration or characteristic mass
A characteristic concentration or characteristic mass describes the amount of analyte that produces a defined analytical response. In absorption methods, a traditional reference point is the concentration corresponding to 1% absorption, which is approximately an absorbance of 0.0044. The source material gives 0.044; that value is inconsistent with the usual conversion for 1% absorption. A corresponding response concept can be applied to emission measurements, though the exact convention depends on the method.
Sensitivity
Sensitivity is the slope of the analytical curve: the change in measured signal for a change in analyte concentration or mass. It can be written as
or, when mass is the calibration variable,
Here, a represents the measured response, such as absorbance or emission intensity; c is concentration; and m is analyte mass. A method has greater sensitivity when a small concentration change produces a relatively large signal change.
Accuracy
Accuracy describes how close a measured result is to the true value. The true value is often not known exactly, so accuracy is assessed using suitable reference materials or other accepted comparisons. A smaller difference between the measured result and the true or accepted reference value indicates better accuracy.
Precision
Precision describes how closely repeated measurements agree with one another under specified conditions. If a sample is measured repeatedly and the results are similar, the method is precise.
Precision does not guarantee accuracy. Repeated measurements can agree closely while all being displaced from the true value because of a systematic error.
The spread of repeated measurements is commonly described by the standard deviation. For n measurements, it is calculated as
where:
xi is an individual measurement,
xˉ is the mean of the measurements,
n is the number of measurements.
A result may be described in relation to its mean and standard deviation, such as the mean plus or minus a stated multiple of the standard deviation. The source material also gives a precision convention involving 2.83s for 30 or more measurements, or ts, where t is a Student’s t factor. The appropriate factor depends on the statistical purpose and method; these expressions should not be treated as universal definitions of precision.
Relative standard deviation
The relative standard deviation (RSD) expresses the standard deviation relative to the mean:
It is often reported as a percentage by multiplying the ratio by 100%. RSD makes it easier to compare measurement variability across results with different mean values.
Determination limit
The determination limit is the lowest concentration that can be quantitatively determined with the precision required by the analytical procedure. It concerns whether a result can be measured with acceptable quantitative reliability, not merely whether a signal can be noticed.
Detection limit
The detection limit is the lowest analyte level that can be distinguished from background under stated measurement conditions. It depends on the background variation and the sensitivity of the method. A general expression is
where:
cDL is the concentration detection limit,
k is a selected factor, commonly 2 or 3,
σ is the absolute standard deviation of the background or blank signal,
S is the calibration-curve slope.
If signal response is expressed as a and concentration as c, then S=ΔcΔa. An equivalent mass-based expression uses the slope relating signal to analyte mass.
Detection limits can be estimated using peak height or peak area. They are affected by:
Electrical noise
Optical noise
Method sensitivity
The analyte’s energy-state behavior, including metastable states
The chemical environment and sample matrix
Thus, a detection limit is not an unchanging property of an element. It depends on the measurement system and analytical conditions.
Detection, identification, and determination
These terms describe different levels of evidence:
Detection: The signal is sufficiently different from background to suggest that the analyte may be present.
Identification: The signal is strong and characteristic enough to support the conclusion that the analyte is present.
Determination: The analyte concentration can be measured quantitatively with the required precision.
A weak signal near the detection threshold may indicate possible presence without supporting confident identification or a reliable concentration measurement. These limits are method-dependent and cannot be represented as universal fixed values.
6. Calibration and quantitative analysis
AAS and ICP-AES are generally relative techniques: quantitative results are obtained by comparing the sample response with responses from reference solutions or reference materials. Reference measurements are therefore necessary.
Calibration must suit the analytical problem. Common approaches include:
Analytical-curve calibration
Known reference solutions are measured to construct a calibration curve. The sample response is then compared with the curve to estimate its concentration.
Bracketing calibration
A sample measurement is bracketed by reference measurements that lie near the expected sample response or concentration. This helps account for changes in instrument response over time.
Analyte-addition calibration
Known amounts of analyte are added to portions of the sample. Comparing the responses before and after addition can help address matrix effects that make external calibration less reliable.
Internal standards
In AES, an internal standard may be added at a known concentration to both samples and calibration solutions. The analyte response is interpreted relative to the internal-standard response, helping compensate for changes such as variation in sample introduction or instrument conditions.
Dynamic range and calibration linearity
The dynamic range of ICP-AES is broad. Calibration graphs may remain essentially linear below about 10 ppm and, for some analytes and conditions, can extend to much higher concentrations, potentially approaching 1000 ppm. When sensitivity is low, a concentrated stock solution may sometimes be measured directly.
This broad range is valuable for simultaneous analysis: it can allow elements present at quite different concentrations to be measured without repeatedly changing the analytical method.
7. Sample preparation, separation, and enrichment
Why sample preparation matters
Sample preparation can introduce contamination, analyte loss, or other errors. This is especially important in trace and ultra-trace analysis, where even a small amount of contamination can substantially affect the result. Whenever possible, unnecessary preparation steps are avoided.
At the same time, samples may need digestion, dissolution, separation, or enrichment before analysis. Each treatment must be selected and carried out carefully.
Separation and enrichment methods
Methods used to separate or enrich analytes include:
Solvent extraction
Activated-charcoal preconcentration
Electrolytic decomposition
Complexation followed by extraction
Chelating agents such as EDTA, ammonium pyrrolidinedithiocarbamate, and NTA can form complexes with metals across suitable pH ranges. The complexes may then be extracted into an organic solvent, such as methyl ethyl ketone (MEK), or another solvent appropriate to the method.
Activated charcoal can adsorb analytes and concentrate them from a larger sample volume. This is a common example of preconcentration.
Digestion and contamination control
Some samples need to be dissolved using acids, sometimes under pressure. Sample treatment can become a source of error if the procedure is not controlled. Ultrapure water and high-purity acids are important in trace analysis, including measurements at parts-per-million and parts-per-trillion levels.
Matrix matching is also important. The reference solutions and samples should be sufficiently similar in composition that differences in signal are attributable to analyte concentration rather than differences in the surrounding chemical environment.
8. Applications of ICP-AES
Biological fluids and tissues
ICP-AES can be used to determine elements such as calcium, magnesium, sodium, potassium, and zinc in body fluids and tissues. Tissue samples generally require wet or dry ashing before analysis.
Other elements of interest in biological samples include arsenic, mercury, cadmium, selenium, and thallium. Aluminum has also received attention in brain-tissue analysis in connection with research on Alzheimer’s disease.
Food and beverages
Food analysis may include animal products such as ghee, butter, and cheese, as well as other foods containing metals. Functional groups and organic material in tissues and food can require digestion, often with nitric acid. Many dairy products are wet-ashed before measurement.
Lead is a common food contaminant and is an important target for monitoring. Tin may also be relevant because it is used in some plastic foils and packaging materials.
Soil, fertilizers, and plants
Elements of interest in soil, fertilizer, and plant analysis include sodium, potassium, calcium, magnesium, copper, manganese, zinc, boron, arsenic, cadmium, and lead. ICP-AES is useful for determining multiple elements in these samples.
Water
Many elements can be determined in water. If analyte concentrations are too low for direct measurement, extraction and concentration procedures may be used. For example, APDC and MIBK may be used in particular methods, although other extraction systems are also available.
Sediment and mineral samples
Sediments and minerals may require aggressive digestion because they can be difficult to dissolve. Hydrofluoric acid (HF) and aqua regia are used in some procedures; digestion choices depend on the material and the elements being measured.
ICP was developed in part for analysis of rocks, minerals, and ores. Common analytes include calcium, magnesium, potassium, iron, copper, zinc, gold, aluminum, and platinum. Digestion using HF and sulfuric acid is one approach used for mineral samples.
Trace-element measurements in geological materials can be supported by ICP-AES and hydride-generation techniques. Volatile elements such as mercury and cadmium may require pressurized digestion, for example in a power bomb. Fusion with reagents such as sodium carbonate or sodium bromide, followed by dissolution of the melt in acid, is another established approach for suitable materials. Gold is often extracted and enriched before determination because its concentration may be too low for direct measurement.
Environmental monitoring
Environmental analysis considers connected systems rather than isolated compartments. Air, water, soil, plants, animals, food, people, waste, and manufactured commodities can exchange substances through physical and chemical processes.
Particulate matter in air has natural and human-made sources. Fine particles around 2.5 micrometers in size can enter and deposit in the lungs. Particles from industrial emissions can travel long distances; the source material notes that stack emissions may be transported as far as roughly 400 km.
Airborne pollutants may be collected by filtering flue dust or other particulates. Sampling periods may be several hours, with durations such as four, six, or twenty-four hours chosen according to the monitoring requirements. Toxic metals, including cadmium, lead, and mercury, may subsequently enter soil, surface water, and other environmental compartments. Sewage sludge and electronic products also require digestion before many elemental analyses.
Metallurgy and plating
In metallurgical materials, elements such as aluminum, vanadium, titanium, chromium, and niobium are commonly measured. Plating solutions may be prepared by treatment with hydrochloric acid and nitric acid.
ICP-AES can determine trace elements such as bismuth, lead, selenium, and tellurium in complex nickel alloys. Aluminum alloys can also be analyzed for elements including cadmium, chromium, copper, iron, and magnesium. One described preparation approach dissolves the alloy in a one-to-one mixture of hydrochloric acid with a few drops of hydrogen peroxide.
Oil, coal, and petrochemistry
ICP-AES is used for trace-element analysis in oil and petroleum products. Iron and copper are monitored in lubricating oils because they may be present as additives; their concentrations may be measured at parts-per-million levels using flame-based methods or ICP-AES. Lead in petrol is another important environmental-analysis target.
Plastics, paper, and textiles
Elements measured in plastics include iron, aluminum, manganese, titanium, silicon, antimony, lead, tin, and zinc. Some polymers, including polypropylene, nylon 66, polystyrene, and PVC, may be dissolved or extracted in suitable solvents before aspiration. Insoluble polymers need digestion or another preparation that brings the analytes into a suitable solution.
Packaging materials may contain organotin compounds used as stabilizers. Metal-containing plastic samples can be extracted with dilute acetic acid; heptane or ethanol extracts may also be measured directly when the method permits. These diverse sample matrices make matrix matching essential for reliable low-level measurements.
Cadmium, copper, lead, and manganese have also been measured directly in paper, pulp, and cellulose using suitable methods.
Pharmaceuticals and miscellaneous products
Pharmaceutical, industrial, and other products may be analyzed for sodium, potassium, calcium, arsenic, cobalt, manganese, silver, and other elements. Some samples are dissolved or leached in concentrated hydrochloric acid before aspiration.
Cobalt in vitamin B12 and proteins can be determined by flame AAS or ICP-AES. Barium may enter products from glass or rubber components as well as through direct ingestion. Dietary supplements are routinely checked for arsenic and selenium.
Uranium in yellowcake can be determined by ICP-AES. Elements such as lead, magnesium, iron, potassium, cadmium, and copper can also be measured in rubber seals and other materials. The broad range of sample types makes method selection and preparation central to successful analysis.
9. Typical ICP-AES analytical workflow
A routine analysis follows a planned sequence:
Prepare the samples and standards. The sample is processed into a form the instrument can measure, and reference solutions are prepared.
Check for an existing method. A validated method may already specify wavelengths, concentration ranges, instrument settings, and other parameters.
Recall and verify the method. If a suitable method exists, it is selected in the instrument software and its parameters are checked.
Calibrate the instrument. Reference solutions are aspirated and measured to establish the calibration response.
Run the samples. The instrument measures the prepared samples using the selected method.
Review and report the results. The results are interpreted in relation to calibration, blanks, quality-control checks, and the analytical requirements.
Developing a method when none exists
When no suitable method is available, the analyst must establish one. Method development includes:
Selecting the elements to measure.
Collecting representative spectra and identifying suitable emission lines.
Selecting the integration time and data-acquisition mode.
Choosing appropriate standard concentrations.
Selecting background-correction positions or procedures.
Storing the validated settings as a method.
Returning to the analysis workflow to calibrate and measure samples with the developed method.
A sequential ICP-AES instrument may be configured as a bench-top system. In a typical arrangement, the plasma is visible within the instrument, while the sample is introduced from below. The sample enters a central region of the plasma, where it is atomized and excited; the resulting element-specific emission is measured for analysis.

Ask anything
We've read this whole study set, so questions about it get the best answers.