MRI Physics Basics
MRI Physics - Lecture 1: Nuclear Magnetic Resonance
🧲 Introduction to MRI Physics
MRI physics is notoriously difficult. The most advanced modern MRI machines are the result of countless specialized subsystems (main magnet, transmit coil, receiver coil), each as complex as a postdoctoral-level subject. However, learning MRI physics does not require understanding every component at once. Just as the first airplane (the Wright Flyer, 1903) obeys the same physics as the F-22 Raptor, the first clinical MRI of a live human (1977) follows the same principles as today's scanners. Building a solid foundation on these core ideas will serve all future understanding.
The three major pillars of MRI physics are:
Nuclear magnetic resonance (NMR) – the phenomenon that makes MRI possible.
Signal localization – how to determine where each signal comes from.
Image contrast – how different tissues appear different in the image.
A unique fourth concept, k-space, connects the MRI equation to image formation and is essential for a complete story.
⚛️ The Proton as a Tiny Magnet
MRI takes advantage of the proton (the hydrogen nucleus) – specifically its charge, spin, and radius. Because a spinning charged particle behaves like a current loop, it creates a magnetic field just like a bar magnet. Therefore, a proton will try to align itself with any external magnetic field, much like a compass aligns with Earth's field.
Key definition: A proton can be modeled as a tiny electromagnet: charge moving in a loop generates a magnetic dipole.
🧭 Net Magnetization
When a human body is placed inside a strong external magnetic field (denoted B₀), the protons inside the body align either with or against the field. Aligning with the field is slightly more energetically favorable, so a very small majority of protons point in the direction of B₀. When summed over billions of protons, this tiny excess creates a net magnetization vector (M) pointing along B₀.
This effect is small (only a few extra protons per million), but because the human body contains an enormous number of protons, the net magnetization is measurable.
Stronger B₀ → greater net magnetization.
Does Net Magnetization Develop Instantly?
No. It grows over time as a simple exponential, characterized by a time constant called T₁. After the body enters the field, the net magnetization increases gradually until it reaches a maximum plateau. T₁ will be explained fully in a future lecture on contrast.
![Diagram of net magnetization growth: a curve rising from zero to a plateau, with time on the x-axis and M_z on the y-axis]
🌀 Larmor Frequency and Precession
If the net magnetization could be pushed away from B₀, it would not stay still. Instead, it would begin to precess (spin around the direction of B₀) at a very specific frequency – the Larmor frequency.
The Larmor frequency is given by a simple formula:
ω0 = Larmor frequency (in MHz)
γ = gyromagnetic ratio – a constant for a given nucleus. For hydrogen (¹H), γ=42.58 MHz/T (often approximated as 42 MHz/T)
B0 = strength of the external magnetic field (in Tesla)
Example: For a 1.5 T clinical magnet:
ω0=1.5×42=63 MHz
This frequency lies in the radiofrequency (RF) range – comparable to an AM radio station.
The Larmor frequency is the natural resonance frequency of the proton system. It is the key to efficient energy transfer.
📡 Exciting the System: The RF Pulse
To push the net magnetization away from B₀, an external oscillating magnetic field (B₁) must be applied. This is done using an electromagnetic coil that generates an alternating magnetic field. If the frequency of this alternating field matches the Larmor frequency (i.e., a radiofrequency pulse), energy is transferred most efficiently to the net magnetization, causing it to tilt away from B₀.
Why Must the RF Pulse Be at the Larmor Frequency?
This is a resonance phenomenon. Just as an opera singer can shatter a wine glass by singing at the glass's natural resonant frequency, an RF pulse at the Larmor frequency causes the net magnetization to precess with maximum amplitude. Other frequencies will have negligible effect.
The flip angle (the angle between the net magnetization and B₀ after the RF pulse) can be set by the strength and duration of the RF pulse.
Common flip angles: 90° (brings magnetization fully into the transverse plane), 180° (inverts it).
Key point: The Earth's magnetic field (≈0.5 Gauss = 5×10⁻⁵ T) is about 30,000–140,000 times weaker than a clinical 1.5–7 T magnet. While MRI is theoretically possible with Earth's field, the signal would be far too weak to be useful.
🔋 Signal Detection: Faraday's Law of Induction
The precessing net magnetization generates a changing magnetic field in the transverse plane (the XY-plane). If a receiver coil is placed nearby, this changing field induces a voltage (and therefore a current) in the coil – exactly the same principle used in wireless phone chargers.
The induced current is sinusoidal and its amplitude depends on the flip angle.
The maximum current occurs when the net magnetization lies completely in the XY-plane (90° flip angle).
The minimum current occurs when the magnetization is aligned with B₀ (0° flip angle).
Additionally, if B₀ (and thus the Larmor frequency) is changed, both the frequency and amplitude of the induced signal change accordingly.
📉 T₂ Decay and Free Induction Decay (FID)
Once the RF pulse is turned off, the net magnetization does not precess forever. Interactions between protons – like friction in a car pushed on flat ground – cause the signal to decay over time. This decay occurs in the XY-plane and is called T₂ decay (or spin-spin relaxation).
Initially, all protons precess in phase (together).
They then interact with each other, some speeding up, some slowing down, causing dephasing.
The net transverse magnetization decreases as dephasing increases, producing a decaying sinusoidal current: the free induction decay (FID).
![Free induction decay curve: a sinusoidal signal with exponentially decreasing amplitude]
The FID as a Sum of Signals
Importantly, the FID recorded by the machine is the sum of the T₂ decay curves from every voxel in the imaged slice. For example, a voxel containing mostly fat has a short T₂ (fast decay), a voxel of soft tissue has a longer T₂, and other voxels have different decay rates. The machine sees only the total sum, not the individual contributions.
The goal of MRI is to deconstruct this complex signal back into the individual decay curves from each voxel. Once separated, a specific time point (called TE – echo time) can be chosen to read the amplitude of each curve. Those amplitudes become the pixel values in the final image – the basis of image contrast.
⏳ The Echo vs. Free Induction Decay
In practice, the initial FID decays too quickly to allow the necessary spatial encoding steps. Therefore, MRI uses techniques to re-create the signal later – this is called an echo. The echo looks like a second, rebounded FID that peaks at time TE and then decays again.
If you cover the first FID portion, an echo looks just like an FID.
The same concepts (amplitude, decay constant) apply to both.
Key insight: All detected MR signal comes only from the XY-plane – the component perpendicular to the receiver coil (by Faraday's law). The Z-component (along B₀) is not detected. This is why T₂ (transverse) decay is directly recorded, while T₁ (longitudinal) recovery is not directly seen – T₁ influences signal indirectly and will be covered in the contrast lecture.
💡 Flip Angle and Practical Limits
There is no theoretical limit to the flip angle – one can apply 60°, 180°, 360°, or even 720° pulses. However, larger flip angles require more RF energy, which can lead to heating of body tissues (specific absorption rate concerns). Therefore, flip angles are chosen carefully.
🔜 Looking Ahead
This lecture covered nuclear magnetic resonance: how protons behave in a magnetic field, how excitation and signal generation work, and how the FID arises. The next lecture will address signal localization – the techniques (frequency encoding, phase encoding) that allow the MRI machine to assign each signal to its correct voxel.
This foundation, though simple, underpins all modern MRI physics. Understanding it thoroughly is the key to mastering every advanced topic that follows.