MRI Physics Basics
- 21Unfamiliar
- 0Learning
- 0Familiar
- 0Mastered
How is a hydrogen proton modeled in the context of MRI physics?
How is a hydrogen proton modeled in the context of MRI physics?
21 questions
- 1
How is a hydrogen proton modeled in the context of MRI physics?
- A
As a current flowing through a loop due to its charge, spin, and radius
- B
As a stationary positive charge
- C
As a magnetic monopole that aligns with external fields
- D
As a neutral particle with no magnetic properties
Correct answer
AAs a current flowing through a loop due to its charge, spin, and radius
Explanation
The proton has charge, spin, and radius, which allows it to be modeled as a current loop, creating a magnetic moment like a small magnet.
- A
- 2
When a human body is placed in an external magnetic field B0, what happens to the protons in the body?
- A
Protons align perpendicular to B0 in equal numbers
- B
More protons align with B0 than against it, creating a net magnetization
- C
All protons align exactly opposite to B0
- D
Protons randomly orient themselves with no net alignment
Correct answer
BMore protons align with B0 than against it, creating a net magnetization
Explanation
The direction aligned with B0 is slightly more energetically favorable, so more protons align with the field, resulting in a net magnetization vector.
- A
- 3
How does the net magnetization develop when a body is placed into B0?
- A
It remains constant from the moment of entry
- B
It develops instantaneously as soon as the body enters the field
- C
It grows exponentially with time constant T1 to a maximum value
- D
It oscillates sinusoidally and then decays
Correct answer
CIt grows exponentially with time constant T1 to a maximum value
Explanation
Net magnetization takes time to reach its maximum, growing as a simple exponential with time constant T1.
- A
- 4
What is the Larmor frequency formula?
- A
ω=γ2B0
- B
ω=B0/γ
- C
ω=γ/B0
- D
ω=γB0
Correct answer
Dω=γB0
Explanation
The Larmor frequency is the product of the gyromagnetic ratio γ and the applied magnetic field strength B0: ω=γB0.
- A
- 5
What is the gyromagnetic ratio for hydrogen protons?
- A
84 MHz/T
- B
42 MHz/T
- C
21 MHz/T
- D
63 MHz/T
Correct answer
B42 MHz/T
Explanation
For hydrogen protons, the gyromagnetic ratio is constant at 42 MHz per Tesla.
- A
- 6
For a 1.5 T MRI magnet, what is the Larmor frequency of hydrogen?
- A
1.5 MHz
- B
42 MHz
- C
105 MHz
- D
63 MHz
Correct answer
D63 MHz
Explanation
1.5T×42MHz/T=63MHz.
- A
- 7
Why must the RF pulse be at the Larmor frequency to excite the protons?
- A
Because higher frequencies would damage tissue
- B
Because lower frequencies cannot penetrate the body
- C
Because the receiver coil only detects that frequency
- D
Because it provides the most efficient energy transfer via resonance
Correct answer
DBecause it provides the most efficient energy transfer via resonance
Explanation
Energy is most efficiently transferred when applied at the natural (Larmor) frequency of the system, similar to an opera singer breaking a glass by matching its resonant frequency.
- A
- 8
How does the Earth's magnetic field compare to a clinical MRI magnetic field?
- A
Earth's field is about 0.5 Gauss, and a clinical MRI is 1.5-7 Tesla, making the MRI about 4 orders of magnitude stronger
- B
Earth's field is negligible, and MRI is only slightly stronger
- C
Earth's field is 5000 Gauss, and MRI is 0.5 Tesla, making Earth stronger
- D
Earth's field is 1 Tesla, and MRI is 1.5-7 Tesla, making them comparable
Correct answer
AEarth's field is about 0.5 Gauss, and a clinical MRI is 1.5-7 Tesla, making the MRI about 4 orders of magnitude stronger
Explanation
1 Tesla = 10,000 Gauss. Earth's field ≈ 0.5 Gauss, clinical MRI ≈ 1.5-7 T, so MRI is about 10,000 to 140,000 times stronger (4 orders of magnitude).
- A
- 9
What is the effect of increasing the flip angle from 0° to 90° on the detected MRI signal?
- A
The signal decreases, reaching minimum at 90°
- B
The signal increases, reaching maximum at 90°
- C
The signal first increases then decreases
- D
The signal remains constant regardless of flip angle
Correct answer
BThe signal increases, reaching maximum at 90°
Explanation
The detected signal comes from the component of magnetization in the XY plane. A 90° flip places all magnetization into the XY plane, giving maximum signal.
- A
- 10
Can the flip angle exceed 90°?
- A
Yes, any angle such as 180°, 360°, or even 720° is possible
- B
Yes, but only up to 180°
- C
No, 90° is the maximum possible
- D
No, because the RF pulse cannot provide enough energy
Correct answer
AYes, any angle such as 180°, 360°, or even 720° is possible
Explanation
There is no limit; we can achieve 180°, 360°, or 720° flips, but more energy is required for larger angles, which affects tissue heating.
- A
- 11
What physical law explains how the MRI machine detects signal from precessing magnetization?
- A
Ampere's law
- B
Ohm's law
- C
Coulomb's law
- D
Faraday's law of magnetic induction
Correct answer
DFaraday's law of magnetic induction
Explanation
A changing magnetic field (from precessing magnetization) induces a voltage in a nearby coil, as described by Faraday's law.
- A
- 12
Which component of the net magnetization vector induces a signal in the receiver coil?
- A
The component in the xy-plane (transverse)
- B
Neither component
- C
The component along the z-axis (longitudinal)
- D
Both components equally
Correct answer
AThe component in the xy-plane (transverse)
Explanation
The receiver coil is oriented perpendicular to B0 (z-axis), so only the transverse component (xy-plane) produces a changing magnetic flux and induces a current.
- A
- 13
What does the free induction decay (FID) represent?
- A
The recovery of signal from an echo
- B
The decay of transverse magnetization due to T2 relaxation after an RF pulse
- C
The growth of longitudinal magnetization after an RF pulse
- D
The net magnetization vector returning to equilibrium along B0
Correct answer
BThe decay of transverse magnetization due to T2 relaxation after an RF pulse
Explanation
The FID is the decaying sinusoidal signal measured after an RF pulse, caused by dephasing of spins (T2 decay).
- A
- 14
What happens to the amplitude and frequency of the MRI signal when the external magnetic field B0 is increased?
- A
Both amplitude and frequency increase
- B
Amplitude increases and frequency decreases
- C
Both amplitude and frequency decrease
- D
Amplitude decreases and frequency increases
Correct answer
ABoth amplitude and frequency increase
Explanation
A higher B0 increases net magnetization (larger amplitude) and increases the Larmor frequency (higher frequency).
- A
- 15
Why is an echo used in MRI instead of relying solely on the free induction decay?
- A
Because echoes provide better contrast
- B
Because the echo signal is purely from T1 relaxation
- C
Because the FID signal is too weak to detect
- D
Because the FID decays too quickly to perform spatial encoding
Correct answer
DBecause the FID decays too quickly to perform spatial encoding
Explanation
The initial FID decays very quickly, so an echo is generated to re-emerge the signal, allowing time for the encoding steps needed to localize the signal.
- A
- 16
The MRI signal from a slice is the sum of what?
- A
The T1 recovery curves from each voxel
- B
The Larmor frequencies from each voxel
- C
The T2 decay curves from each voxel
- D
The RF pulses applied to the slice
Correct answer
CThe T2 decay curves from each voxel
Explanation
The measured signal is the sum of all individual T2 decay curves coming from every voxel in the slice.
- A
- 17
How is an image formed from the raw MRI signal?
- A
By measuring the average signal from the entire slice
- B
By applying a Fourier transform to the raw signal without additional steps
- C
By decomposing the sum signal into individual voxel decay curves and sampling them at a specific time TE
- D
By directly displaying the raw signal as an image
Correct answer
CBy decomposing the sum signal into individual voxel decay curves and sampling them at a specific time TE
Explanation
The goal is to separate the combined signal into contributions from each voxel. The amplitude of each voxel's T2 decay at time TE is used to assign pixel brightness.
- A
- 18
What is the role of the echo time (TE) in image formation?
- A
It is the time at which the amplitudes of individual T2 decay curves are sampled to create contrast
- B
It is the repetition time between successive excitations
- C
It is the duration of the RF pulse
- D
It is the time when the RF pulse is applied
Correct answer
AIt is the time at which the amplitudes of individual T2 decay curves are sampled to create contrast
Explanation
At TE, the signal from each voxel's T2 decay curve is measured, and these values determine the contrast in the final image.
- A
- 19
Which relaxation process is responsible for the decay of the transverse magnetization signal?
- A
T2 relaxation
- B
T1 relaxation
- C
Spin-lattice relaxation
- D
Both T1 and T2 equally
Correct answer
AT2 relaxation
Explanation
T2 relaxation (transverse relaxation) causes dephasing and decay of the signal in the xy-plane.
- A
- 20
What is the net magnetization vector after equilibrium in B0?
- A
It rotates in the xy-plane at the Larmor frequency
- B
It is zero because individual moments cancel
- C
It points opposite to B0
- D
It points along B0 (longitudinal direction)
Correct answer
DIt points along B0 (longitudinal direction)
Explanation
At equilibrium, the net magnetization is aligned with B0 (along the z-axis).
- A
- 21
What happens to the net magnetization vector when an RF pulse at the Larmor frequency is applied?
- A
It aligns more strongly with B0
- B
It disappears
- C
It stops precessing
- D
It is tipped away from B0 and begins to precess
Correct answer
DIt is tipped away from B0 and begins to precess
Explanation
The RF pulse transfers energy to the system, tipping the net magnetization away from B0; it then precesses around B0 at the Larmor frequency.
- A

Ask anything
We've read this whole study set, so questions about it get the best answers.