MRI Physics Basics
- 18Unfamiliar
- 0Learning
- 0Familiar
- 0Mastered
Why does a hydrogen proton behave like a small magnet?
18 questions
- 1
Why does a hydrogen proton behave like a small magnet?
Model answer
A hydrogen proton has a positive charge, spin, and radius, so it can be modeled as a current flowing through a loop, which generates a magnetic field.
Explanation
The proton's charge and spin create a circulating current, analogous to an electromagnet.
- 2
When placed in an external magnetic field, do more protons align with or against the field, and why?
Model answer
Slightly more protons align with the magnetic field because that direction is energetically more favorable.
Explanation
The energy difference between alignment states is small, but the excess population with the field creates a net magnetization.
- 3
What is the net magnetization vector (M) and how is it created?
Model answer
The net magnetization vector is the sum vector of all proton magnetic moments, resulting from a slight excess of protons aligned with the external field.
Explanation
This net vector points along the main magnetic field (B0) and is the source of the MRI signal when manipulated.
- 4
How does the net magnetization develop over time after a body is placed in a magnetic field?
Model answer
It grows exponentially with a time constant called T1, reaching a maximum value.
Explanation
T1 represents the time for longitudinal magnetization to recover or establish itself after initial exposure.
- 5
Write the Larmor frequency formula and state the gyromagnetic ratio for hydrogen.
Model answer
The Larmor frequency is f=γB0, where γ=42MHz/T for hydrogen protons.
Explanation
This formula is fundamental to MRI, as it determines the resonance frequency of protons in a given magnetic field.
- 6
Calculate the Larmor frequency for a 1.5 Tesla MRI magnet.
Model answer
The Larmor frequency is 1.5T×42MHz/T=63MHz.
Explanation
This frequency is in the radiofrequency range, which is why RF pulses are used for excitation.
- 7
Why must an RF pulse be applied at the Larmor frequency to excite protons?
Model answer
At the Larmor frequency, energy is transferred most efficiently to the spin system, similar to resonance in a mechanical system.
Explanation
Just as an opera singer matches the natural frequency of a glass to break it, RF at the Larmor frequency maximizes energy absorption and causes precession.
- 8
Why can't the Earth's magnetic field be used for clinical MRI?
Model answer
The Earth's magnetic field is about 0.5 Gauss, while clinical MRI uses 1.5-7 Tesla (1 T = 10,000 G), making Earth's field about four orders of magnitude weaker.
Explanation
A stronger field is needed to create a significant net magnetization and sufficient signal.
- 9
What happens to the net magnetization when an RF pulse at the Larmor frequency is applied?
Model answer
The net magnetization is tipped away from the main field and begins to precess at the Larmor frequency.
Explanation
This precession is the basis for generating an MRI signal through Faraday induction.
- 10
How does the MRI machine detect a signal from the precessing net magnetization?
Model answer
The precessing magnetization creates an alternating magnetic field that induces a voltage and current in a receiver coil via Faraday's law of induction.
Explanation
This is analogous to a wireless charger, but much more sophisticated.
- 11
Why does the MRI signal come only from the XY plane?
Model answer
Because Faraday's law requires a changing magnetic flux perpendicular to the coil; the component of magnetization along the z-axis does not induce current.
Explanation
Only the transversal (XY) component of the magnetization produces a signal in the receiver coil.
- 12
What is the free induction decay (FID) signal?
Model answer
It is a decaying sinusoidal current induced by the precessing net magnetization as it dephases due to T2 decay.
Explanation
The FID is the raw signal immediately after an RF pulse, decaying with time constant T2.
- 13
What causes T2 decay of the MRI signal?
Model answer
T2 decay results from interactions among protons that cause them to lose phase coherence (dephasing), reducing the net transversal magnetization.
Explanation
Protons speed up or slow down relative to each other, causing the signal to decay.
- 14
What does the raw MRI signal represent?
Model answer
It is the sum of T2 decay curves from every voxel in the selected slice.
Explanation
Each tissue type has its own T2 curve, and the machine records the combined signal.
- 15
What is the goal of further processing the raw MRI signal?
Model answer
To deconstruct the complex raw signal into individual T2 decay curves from each voxel, allowing image reconstruction.
Explanation
By separating the signals, we can assign intensities to each voxel and create an image.
- 16
Why is an echo used instead of the initial FID for signal localization?
Model answer
The initial FID decays too quickly to perform the necessary spatial encoding steps, so the echo re-emerges the signal for measurement.
Explanation
The echo provides a delayed, refocused signal that is more practical for building an image.
- 17
Is there any limit to the flip angle that can be achieved with an RF pulse?
Model answer
No, any flip angle can be achieved (e.g., 60°, 180°, 360°, 720°), but larger angles require more energy.
Explanation
The energy of the RF pulse determines the flip angle, with clinical implications for heating.
- 18
What does T1 represent in MRI physics?
Model answer
T1 is the time constant for the longitudinal magnetization to recover after excitation, growing exponentially back to equilibrium.
Explanation
T1 contrast is important for tissue differentiation, though detailed coverage is deferred.