boom all right so again thank you guys for stopping by today and um taking part in our little i just say scientific discussion on gas chromatography mass spectrometry as many of you probably know the science department does have a gc mass spec located in ms 212 and you know this workshop presentation is mostly to give you some idea of the things that we have been doing as far as using the gc mass spec and then hopefully uh prop some ideas promote some ideas for your classes to possibly use the gc mass spec as well alright so i know the title is a little longer than what was in the initial description the actual title was design and implementation of practical qualitative experiments involving gas chromatography mass spectrometry as an analytical tool and again if you guys are just joining please raise your hand during the discussion and you can open mic and ask a question all right so to begin
we just want to become familiar first with gas chromatography mass spectrometry the instrumentation and operation i've got a background in mass spectrometry and so there are a few more details that i'll share i won't go too far in depth i hope but the idea is to let you know what the instrument is capable of doing then we want to discuss the types of samples sample amount and sample preparation that might seem trivial but this instrument is very sensitive i can pick up fiptomolar and that's femto meaning the prefix femto molar amounts of sample and so it's very critical that you prepare your sample properly when you get ready to use the gc mass spec and then finally we got a showcase at the end we have two projects one involving uh cbd oil that we've done with my class and then one that david will cover we talk about retention bonds
all right so let's go ahead and talk about what gc mass spec is it's an analytical method that combines the features of gas chromatography and mass spectrometry to identify different substances within the test sample and i point out that it is an analytical method of course everybody knows i guess there are five areas of uh chemistry analytical chemistry being one specific area i'm an analytical chemist i know dr carlos baichi diaz is also an analytical chemist and in analytical chemistry we study and use instruments and methods to separate identify and quantify matter and so you might think of this notion of separating and identifying if you're identifying things that's the qualitative part okay and so that's what our entire field does so this instrument is primarily used in analytical chemistry of course it's two things it's gas chromatography that's a common type of chromatography used in analytical chemistry
for separating and analyzing compounds that can be vaporized without decomposition as the name states it's gas chromatography all right so you've got solid liquid and gas in this case if the sample cannot be turned into a gas at a regular temperature meaning up to 120 to 300 degrees then you probably can't use it in this instrument and then lastly here we have mass spectrometry is an analytical technique that's used to detect and measure mass to charge ratio of ions so
here's a little summary of kind of that same thing here so gas chromatography mass spectrometry is what we call a hyphenated analytical technique which means it's two techniques that are combined to form one single method of course the gc again is dealing with volatile organic compounds all right and the mass spec is our detector so let's get into a more detail of the instrumentation this is a picture of the instrument that we have in ms212 there are a number of parts up here that you'll see in another diagram coming up but i just wanted to point them out so in case you come into the room you'll know what the system is designed for and have all the parts for it i got a carrier gas over here in the corner so that carrier gas is a helium gas for this particular system got an injection port and this injection board i wish i had a video of it but anytime you watch a csi program on television they'll typically have the auto sampler for the gc running because it
spins around it's a little robot arm that picks up the sample and drops it into a area so that you can have a needle pierce it and then inject into the actual instrument and then finally have this region the gc itself then your mass spectrometer and your data system so let's dig into these with a little more detail the first thing we want to talk about is the carrier gas okay so when we talk about chromatography now you got to talk about two different phases you have what we call a mobile phase which as the name is stated it's a moving phase and so in the case of our carrier gas which sits in our tank the carrier gas pushes in our case helium throughout the entire system so the carrier gas is hooked up to the injector the injectors hooked up to this gas chromatography column and then your column goes into your mass spectrometer so the carrier gases that we usually use are going to be inert gases meaning that
they are not reactive gases if you put them with other compounds or chemicals they don't cause a reaction with those other chemicals so we're talking about things like helium and nitrogen now on occasion they'll use hydrogen sometimes hydrogen's a little bit cheaper but in our particular case we don't want to use hydrogen because hydrogen is flammable so our carrier gas is running through the system the next area you'll note is the injector port okay so the injector point is that region that i pointed out that was on top of the mass spec in our diagram it's also got a syringe and the injector injects the sample into the flow of the carrier gas and in this region your carrier gas passes through onto your column once the sample gets in it's immediately volatile so it's turned into a gas by a high temperature typically our injection injection ports are sitting at about 120 degrees celsius
so if you're familiar with the boiling point of water you know that water boils at 100 degrees most of the samples that we want to inject can be turned into a gas at much lower temperatures so if you enter that sample put that sample into this injection port at 120 degrees or 150 degrees in some instances automatically the sample gets turned into a gas so samples are can be diluted we talked about the sensitivity how little amount of sample that you have to be able to put into this instrument we can dilute the sample at the injection port okay so that's one of the details that makes this instrument pretty nice is that you can do some dilution within the system itself
next we're going to go to the gc column and the column itself is placed in an oven it literally is an oven and the oven of course can go from 40 degrees all the way up to 350 degrees celsius and the idea is that you are slowly heating the column over time and as you slowly heat the column samples that are slow to be moving because they're at a larger molecular weight where they have more
interaction with the column itself they actually start to boil and move more quickly through the column okay so the whole separation process takes place as a result of heat and the column itself the columns are generally about 15 meters to 60 meters long so that's that coil that you see here in my diagram and the key there is that the length of the column gives you theoretical plates that's a principle but the short end of it is you can think of it as rocks in a river
so the gas chromatography or the carrier gas is like the water flowing through the river and then if you have rocks in a river and you have a small raft you can easily navigate the river but if you have a large raft then it becomes more difficult to navigate all the rocks and therefore you move more slowly so certain compounds will interact with the gc column more than other ones and if they interact more as a result of the column and the compound itself then they'll come out of the column last okay so columns are a stationary phase that will retain compounds based on how they interact with the column this is a side note when i worked in industry
the guys in industry talked about how they used to take their cans of soup before they had microwaves and they would sit them in the gc ovens and they would heat their cans of soup inside the gc ovens of course we wouldn't do that today there'd be a safety issue with that but it is just another all right so the next system there is going to be your mass spectrometer and i'm sorry one more detail about the column so your things are getting separated on the column and as they're coming off the column you can note here this is a counts versus acquisition time so the first items that you see down here about 4.4 4.6 at about 4.5 this compound came off of the column first and then you can see successively that each of these other compounds comes off the column at a later time the idea being that these columns that come off very late
in this case around 7.8 minutes those columns had a lot of interaction with are those compounds had a lot of interaction with the column and so they came off later and this becomes important because this retention time is one of the things you can use to help identify your sample all right so now on to the mass spec uh so your mass spec is your detector all right so compounds are separated as transferring through the heated area right here between the two instruments and then you have three parts to the mass spectrometer i have an ionizer a mass filter and a detector let me go into that in a little more detail so over here you have a gc stream
and your gc compounds the compounds that are being separated are going to flow through the entrance of the source so this is the ionizer this area right over here notice that i've got an arrow here to neutral compounds so things that come through the gc have no charge meaning they're not positively charged or negatively charged they're neutral when they get into the ionizer they go through a beam of electrons
and these electrons interact with each one of these compounds and they knock off an electron and if you remember from your general chemistry if you remove an electron from a compound ultimately create a positive charge because the electron is negatively charged so by passing through this beam now each one of these compounds has a positive charge so that's the ionizer that's what it does is create ions from there it passes through what we call a mass filter and there are different types of mass filters but this one that i have represented here is called a quadrupole and it's essentially four different rods and in these four different rods you will create or you'll have different voltages so the top one and the bottom one might be positively charged the one at the back and the one and closer to the front would be negatively charged and then you alternate those voltages very quickly and what that does is if you have something that's
positively charged moving through the rods as they encounter the positive rod they'll start to move away but then as they start to go up and you flip the voltage now that positive charge is on the top right and they'll be pushed back down so you essentially filter the positive charges through this region all right so that's why it's the mass analyzer now we place a dc voltage on this and we scan it and that dc voltage allows us to get what we call the mass to charge ratio um
and basically a massive charge ratio if you have a compound like water and water is a mass of 18 then the charge on it would be one so 18 divided by one would be one so the mass that you would see is 18. all right so once it gets through there then it goes on to your detector and once it gets in the detector it strikes a plate that causes an electrical signal all right so that's the details of our mass spec that electrical signal goes over to the data system
and our data system has basically two things that it can show obviously you've got your chromatogram that was based on your gc so you've got all your various compounds that are coming out and then you've also got a mass spectrum and in this case a mass spectrum with your master charge ratio gives you different pieces or different ions that are based on the one compound that came through the instrument at a single time so if we go back to the compound that came out at position or time number four on this particular chromatogram
all of this mass spectrum represent that one compound at that one particular time and we'll get into more detail about that in just a moment all right so in summary you've got to carry a gas you inject your sample the gas chromatograph actually separates the mixture your mass spectrometer detects the compounds and then your data system collects and analyzes your data all right so that's it for the first part i hope that was good there as far as explaining kind of the components of the gc mass spec let me move on to some of the sample preparation we'll talk about sample state and amount first and the key there i said earlier is that you have to have compounds that are going to be volatile so compounds must be soluble in a solution for them to be injected onto the gc the solvent itself has to be volatile
generally organic so for example we will use hexane and we'll use dichloromethane uh hexane i if you guys will note is a nonpolar solvent while dichloromethane is a polar solvent all right the key is here that however water being polar water should not be used as a solvent water is a no no in gc and that's because if you put water in the instrument water's boiling point is 100 degrees and if you were to inject it on the gc column it would
actually crash out as a liquid and that would be a liquid until it reached the temperature above 100 degrees at which point it would start to turn into a gas and then essentially it would fill up the entire column because you have so much of it okay so water we don't want to use the hexane and the dichloromethane boil in much lower temperatures and as a result they'll be off of the column very quickly because they don't need a high temperature for them to go and turn into a gas
for the amount of sample that's going to depend on the ionization method in our case it's electron impact but the sensitivity of the instrument is about one to one hundred picograms per component and so i always tell dr vajradias with his organic class dilute it once then dilute it again and finally dilute it in the instrument because the instrument is really very very sensitive the injection amount is actually one microliter and they give you some idea of one microliter injection for that liquid if you inject or push the syringe down the one microliter does not have enough weight to actually drop off the end of the syringe okay that's how small of an amount that that one microliter is okay so let's move on now to the sample preparation so sample preparation can change from simply dissolving some of the sample in a suitable solvent to extensive methods and we'll kind of see that today with the two
projects that we'll show the first thing here with your cleanup procedures they can take various forms you can have liquid liquid extraction so if you look over here to the side this might be a simple liquid liquid extraction in this case you might have hexane on top and then water on the bottom and the water would be drawn out of the bottom of the container and then the hexane would be drawn out and then you'd re-add your water and then your hexane add a new layer of hexane and you shake this up to actually extract your compounds from your water that's a liquid liquid extraction or we've also tried solid phase extraction in a couple of places i know we tried them in geology and solid phase extraction you have a solid phase that's located in a typically a tube and you add your sample to the tube and then using pressures either negative or positive pressure
you allow the sample to pass through a solid phase i always tell people it looks like sand but it's actually silica glass that's been modified so you can capture quite a bit of your matrix the key here is that you want to take as much of the dirty stuff so for example if you literally had dirt you could mix it up pour it on top and then the dirt would stay in the tube and your soluble components would pass through the syringe area and you can collect them so you want to get rid of the matrix and then put as much of the compound of interest into the instrument so those are just two methods there are quite a number but these are pretty common methods all right so how do we perform this qualitative analysis uh so the first thing here this is different from quantitative okay i want to point that out early on if you uh look at qualitative data
qualitative data just gives you information about the properties or the attributes of your sample so in this picture you have different colors you have a precipitant this can tell you that a compound is there but it does not tell you quantitative data which is how much of a sample you have so maybe in the future we'll give a presentation about doing actual quantitative analysis with the gc mass spec but we're talking about qualitative we generally just want to figure out what compounds are there
so qualitative analysis uh your chromatography this is all from your data system chromatography gives the retention time for your compound so in this particular case this is a chem 151 sample they are asked to make aspirin from a salicylic acid and in this case the student did not do so well the salicylic acid can be found here at 8.35 and you can look at the mass spectrum this fragmentation pattern that you have over here tells you that you have salicylic acid all right so those are the two main ways that you'll identify the compound the first is the retention time and then the other one can be the mass spec pattern
so for your gc if you have a compound for example the salicylic acid you can't really identify the compound without the mass spec and so what you'll do is you'll inject a standard and so i would go to the lab and i would get a standard of salicylic acid and i would inject it using the same chromatography conditions and it would give me a retention time and then when i inject a sample of unknown if i have a compound that comes out at the same retention time then i can probably say that that same compound is the standard or is the same as the standard in this case it would be salicylic acid okay so as long as the gc method is unchanged compounds will elute from the gc column at the same or very similar retention times so that's why we use our standard as far as the mass spec is concerned we can use a database so we'll go over to what we call the webbook.gov.nist that's the website
and you can type in salicylic acid and when you type that in it'll give you this page and right down here it'll say the mass spectrum for it so you can look at that mass spectrum and match it to your mass spectrum and gc fragmentation is very specific it creates the same fragmentation pattern for that compound each time and so we can match this fragmentation pattern to the fragmentation pattern in the database and that will tell us that we have that compound
so here's that example this one's from the nist web book so you see here the 138 120 92 for the salicylic acid in our gc sample you see the 138 the 120 and the 92 and the abundances meaning the intensities are about the same so that tells you that you probably have the salicylic acid right if you do not have a database then you have to interpret the mass spectra or have a known standard of the compound uh if you want uh we can do some discussion later about how to
interpret mass spectra but i can tell you that's an entire course within itself all right so let's move on now to the qualitative analysis we've got two different projects that we've been working on the first one here is qualitative analysis of cannabinoid which is also known as cbd and then i'll let david introduce his a little bit later that's the qualitative analysis of vocs in retention pond water so first thing about cbd i think most people have heard of cbd if you go to the grocery store i know in the checkout lines at the grocery store they have cbd products
i'm pretty sure they're sold in the malls i'm pretty sure they're sold at the vitamin shop the gnc you can order them from all over the place so it's become a very hot topic gcml's gcms analysis can be used to look at the components found in any of these cbd samples and chemical structure is one of the objectives for general chemistry and so i have over here to the side two different cbds now if you look up cbd you don't find these two structures this is actually from the nist website and for those who are savvy they notice that everything else in these two structures is the same except for this little wedge that's on the top here this wedge is solid and down here on the lower one it's a dashed wedge so if you took dr vajidi as his organic class you would note that the wedge that's solid means that this front part is coming towards you and that's a chiral center
and on this lower one the wedge or the dashed line means that this part is going away from you and that's another chiral center okay so it looks like there are two different cbd components so in this case we'll look at this in a second but uh we want to pay attention to that please remember that you have two different cbd's there so the sample preparation is just dilution in this case you have a chromatogram of our cbd samples we have a standard that came off the column at about 14.54 minutes and so this is a known standard we just diluted it and put it directly in the system to check and make sure that we had retention time and that it was what we say it is the actual cbd results for the cbd compound show a few other components right so this particular sample on the bottom is a sample that we ordered and this sample has more peaks in it than the standard which it should right because cbd is a natural product
so the additional peaks that were found in the sample we tried to identify those using mass spec and what we find is that the peak that's at the back is thc now i point that out because i'll show you here in a moment that um the fda has a level uh cannabinoid is not i'm sorry the cannabis plant the hemp plant is not a marijuana plant and so therefore it has a level of thc in it but you should not have a huge amount of thc in it you also point out that these
other two peaks this is the peak at 14.54 that is the cpd standard but i found another peak at 13.96 that's also a cbd so if you look this up in the database it has a very similar mass spectrum to the other mass spectrum for the cbd so there's the instrument itself seems to have resolved two different cbds but there's only one that we ordered for the standard okay so without the mass spec we wouldn't know what these other components are okay and just as a side note uh just because we're dealing with cbd here i'm not going to read all this to you guys but i do want to point out that cbd is a single component found in the cannabis plant and because it is these other components are probably there so when students get in my classroom and they start to talk these things i point out to them that there are a ton of other things kind of like the tea tree oil
had a bunch of stuff in it there can be a number of other things that are found in these things the fda has only approved one cbd product and that's to treat things like seizures these other type of things up here i know the seizures are a big deal you can actually go out and google cbd pbs and they have quite a number of videos that talk about how cbd is used to treat seizures and then there are a couple of other details down here but the key is that you know cbd itself is a product where it's one compound but a lot of the things that students are talking about have a number of other things that can be shown with gc mass spec all right so let me switch over uh david did you want to
fill us in on your background for your retention pond water yeah so i had this idea a while back um because in my geology 111 class in lab i do a flood hazards lab and if you weren't aware of these ponds that we have on campus and all around eastern north carolina are a very simple way to control flooding so what they do is they storm they store the storm water temporarily and they let it the the water soak into the ground more naturally versus uh storm drains that that pipe that directly to the rivers and that can cause flash flooding so i had this idea that since we're we're we're draining this impervious surface aka parking lots where we can have vehicles that uh leak all sorts of petroleum that that possibly over time that the stormwater retention ponds
could be point sources for pollution means we know exactly where that pollution came from so we've got some uh results of of uh that we if we spike the sample we can retrieve it um but we have not yet been able to detect it in the water now you can see in the upper right hand corner of the screen there you can see this is this is right outside the where i park every day and this was right after a rain so that is oil that is on the ground and so you can see that that is going to eventually end up into the retention pond uh depending on the results of the water i would also like to do some coring of soil around the ponds to see if we can detect any uh petroleum in the in the soil as well
but because of laboratory contamination we're having some issues with with getting some good results uh hopefully at some point in time we'll get this instrument called a purge and trap where it does the extraction for you and injects into the gc mass spec so the the points of contamination is dramatically reduced but as of yet we've not had we've not been able to uh acquire that piece of equipment and that's all all right thank you david and like like you said i
mean uh we've got a project or a method for this and that method is very involved uh that's kind of why we want to try and get the purging trapped um the method itself is a liquid liquid extraction and this is just a diagram of liquid liquid extraction i kind of talked about it earlier um but if you've been to the chemistry lab in room ms212 and even in the prep room in between you've probably seen these huge two liter uh separatory funnels and what we do there is we'll add in about 500 milliliters of pond or retention pond water and then we'll add in maybe 30 to 100 milliliters of dichloromethane or hexane and you shake your sample with those particular solvents and then of course the dichloromethane is more dense so it would be on the bottom hexane would be on the top
but either way you let the elements out of your separatory funnel and then you might do two or three other extractions to try and get your sample cleaned up and the key here is that you want to concentrate your sample because the pond water is going to have such a small amount of compound in it it's important to take almost a liter of sample and concentrate it down to maybe a milliliter so you're doing a thousand whole concentration there these are some of the details of what we're using so we use a deuterated standard for analysis and when you hear the word deuterated that means you have three different isotopes for hydrogen isotopes meaning that you have more neutrons it's the same element but you have more neutrons in the nucleus and a neutron is going to make it way a little bit more it's going to make it way one atomic mass unit more and so we have a deuterated standard of octane because of course we'd be looking for
things that are gasoline type so octane itself and we pass it through that liquid liquid extraction procedure the key is that we can spike that octane standard directly into our water sample before we start to do the extraction process all right so results here are just from the mass spec your standard and then the mass spectra from that particular compound so here's the compound itself and what you want to think about is for every single hydrogen that you would have on the octane
you're going to get one deuterium and so that deuterium is going to add one additional mass unit so what that means is that you can put this particular compound into the extraction and even if octane is in the extraction or the pond water itself then you can still have them separated by mass to charge the mass spec can tell the difference the gc cannot tell the difference but the mass spec can tell the difference the retention times for the two compounds would be the same but the mass spec tells you it's different and the big thing there is that later on we'll use this standard for qualitative or quantitative analysis because we can take peak areas and we can normalize things because we know how much of this particular compound that we are adding to the sample and so we can use that to figure out how much we've recovered when we take it through the extraction process
all right so that's our pond water analysis these are other applications up here um that are just from literature and hopefully you'll see one up there for you um let me encourage everybody to you know if you want to use the gc mass spec for any of your laboratory work uh please let me know there's probably a technique out there that we can use even if it's plants i know i've worked with mr horn in the past on trying to do things for tomatoes we've also had a number of other things like i say the aspirin lab if you want to do kinetics we can do things like that so with that i think that concludes everything for us does anyone have any questions please
it's quiet as a mouse okay all right well if you guys have no questions i know i got a question yes sir uh if could you use this technique to measure ethanol concentrations like produced by fermentation yes sir you can in fact that would be a great project because that's just a specific gc there's an actual gc method for looking at the various alcohols and so we could just switch over to that method make the injection and go from there but yes that can be done very easily all right thanks
other questions i see carlos has one up here you can also use an ir instrument to do that yes he's right you can also use the ir instrument to do that and we also have the lc instrument that you might be able to do that with as well so there are a number of analytical pieces of equipment mark that we might maybe can use to do that analysis you spoke of how gc fragmentation is very specific however and i'm trying to word my question as best i can based upon my understanding of your very well very good presentation are there any redundancies among chemicals
uh so are you saying are there fragmentation patterns that are similar yes among compounds yes in fact i'm sorry even with the example you presented um yeah i think you you interpreted my question correctly yes yes so yeah the uh compounds themselves the two that i presented have chiral centers so those were resolved because we had the gc in front of them however without the gc the mass spec could not tell the difference and so that's the key is that the mass spectral mass spectrum for each of those compounds would look the same
but we know that they're different because they had different elution time illusion times on the gc so you are correct that you'll have very similar patterns and i see uh dr carlos actually pointed out that hydrocarbons have very similar patterns they do hydrocarbons break apart um based on the number of carbons that they have so for example if you had octane you can expect to see between eight and probably six groups that are separated by 14 daltons in the mass spec the issue becomes that there are different arrangements of those hydrocarbons so it doesn't have to be a long chain it can actually have side chains and things like that generally side chains will break off and you'll have a very significant gap when you actually interpret the mass spectrum
but you are correct that you can have some and we use those groups those very similarities in the mass spectrum to say that yes you do have a six-membered ring for example you have a benzene ring fragment that's very well known or if you break off a hydroxyl which means you have a loss of 18 because you have hydrogen and oxygen and then you add one hydrogen to it so you can lose 18 that specific loss tells you that you have a certain pattern and so again it would takes a little bit to be able to interpret mass spectrum right there's a lot of books on how to interpret mass spectrum but they do in some cases have a lot of overlap okay great thank you
and i see carlos's point up here you know that's a fascinating thing carlos i would like to get some understanding of what's going on in that gc mass spectrum it was identified carlos is saying that those enantiomers the two different cbds have very similar structures so theoretically we were not doing chiral chromatography so the only thing that would separate them would be their boiling point or their interaction with the column and so that's a real question that i have is you know was that first peak it showed up as cbd i need another standard of a different chiral center cbd to say for sure that it is a different cbd that's a good point yet but diastone would have a different boiling point that's correct
so those are the details in there that and i thank you um dr value for your interjection because i think that's where my thought was going when i asked about the redundancies i just didn't know to relate it to the enantiomers or the diastereomers so thank you yep good science are there other questions any discussion points you want to add anybody want to add anything
thanks carlos all right well if no one has anything else let me encourage you guys you know if you think of anything like mark was saying look at fermentation or even if you want to look at constituents of some plant that's in the greenhouse or anything like that or if you're doing i know we're not supposed to use human blood or if something like that comes up we can try and always run them on the mass spec and think of some different projects to be used on the mass spec so please let me know if you have any uh ideas and i'll be around thank you guys for coming by today i appreciate it